Table C.8.31.5-1 specifies the Attributes of the X-Ray Acquisition Dose Functional Group Macro.
Table C.8.31.5-1. Breast X-Ray Acquisition Dose Macro Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
X-Ray Acquisition Dose Sequence |
(0018,9542) |
1 |
Sequence containing the X-Ray exposure information for this frame. Only a single Item shall be included in this sequence. |
|
>Exposure Time in ms |
(0018,9328) |
1 |
Duration of X-Ray exposure in milliseconds. |
|
>Exposure in mAs |
(0018,9332) |
1 |
The exposure expressed in mAs, for example calculated from Exposure Time and X-Ray Tube Current. |
|
>Relative X-Ray Exposure |
(0018,1405) |
3 |
Indication of the applied dose, in manufacturer specific units. Note
|
|
>Half Value Layer |
(0040,0314) |
3 |
The thickness of Aluminum in mm required to reduce the X-Ray Output (0040,0312) by a factor of two. NoteThis value may be a calibrated value rather than measured during the exposure. |
|
>Organ Dose |
(0040,0316) |
1 |
Average organ dose value measured in dGy during the acquisition of this image. NoteThis may be an estimated value. |
|
>Entrance Dose in mGy |
(0040,8302) |
1 |
Average entrance dose value measured in mGy at the surface of the patient during the acquisition of this image. NoteThis may be an estimated value based on assumptions about the patient's body size and habitus. |