Table C.8.31.4-1 specifies the Attributes of the Breast X-Ray Geometry Functional Group Macro.
Table C.8.31.4-1. Breast X-Ray Geometry Macro Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
X-Ray Geometry Sequence |
(0018,9476) |
1 |
Sequence containing the geometric properties for this frame or set of frames. Only a single Item shall be included in this sequence. |
|
>Distance Source to Detector |
(0018,1110) |
1C |
Distance in mm from source to detector center on the chest wall line. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. Note
|
|
>Distance Source to Patient |
(0018,1111) |
1C |
Distance in mm from source to the Breast Support side that is closest to the Imaging Subject, as measured along the X-Ray beam vector. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. Note
|
|
>Distance Source to Isocenter |
(0018,9402) |
1C |
Distance from source to isocenter in mm. Required if Presentation Intent Type (0008,0068) = FOR PROCESSING. May be present otherwise. |
|
>Distance Source to Entrance |
(0040,0306) |
3 |
Distance in mm from the source to the surface of the patient closest to the source during the acquisition of this image. NoteThis may be an estimated value based on assumptions about the patient's body size and habitus. |
|
>Estimated Radiographic Magnification Factor |
(0018,1114) |
1 |
Ratio of Source Image Distance (SID) over Source Object Distance (SOD). |