The templates that comprise the Quantitative Ventricular Analysis SR are interconnected as in Figure A-3:
The Ventricular Analysis Template provides a CONTAINER with a structure for reporting the result of the ventricular analysis.
Extensible
Significant
Table TID 3202. Ventricular Analysis
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
|
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1-n |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
5 |
> |
CONTAINS |
CONTAINER |
1-n |
M |
|||
|
6 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1-n |
U |
|||
|
7 |
>> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
8 |
>> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
9 |
>> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
10 |
>> |
CONTAINS |
IMAGE |
1-n |
M |
|||
|
11 |
>>> |
HAS CONCEPT MOD |
CODE |
EV (G-A60B, SRT, "Cardiac Phase") |
1 |
M |
||
|
12 |
>>> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
If Biplane Analysis |
||
|
13 |
>> |
HAS ACQ CONTEXT |
INCLUDE |
1-2 |
U |
VM = 1: Single plane analysis, VM = 2: Biplane analysis |
$CalibrationPlane = DCID 3466 “Plane Identification” |
|
|
14 |
>> |
HAS ACQ CONTEXT |
INCLUDE |
1 |
U |
|||
|
15 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
16 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
17 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
18 |
>> |
CONTAINS |
INCLUDE |
1-2 |
U |
VM = 1: Single plane analysis, VM = 2: Biplane analysis |
||
|
19 |
>> |
CONTAINS |
INCLUDE |
1-2 |
U |
VM = 1: Single plane analysis, VM = 2: Biplane analysis |
||
|
20 |
>> |
CONTAINS |
INCLUDE |
1-2 |
U |
VM = 1: Single plane analysis, VM = 2: Biplane analysis |
Content Item Descriptions
|
Row 7 |
Identifies the Ventricular Analysis program |
|
Row 8 |
Identifies the Ventricular Analysis program version |
|
Row 9 |
Identifies the Ventricular Analysis program manufacturer |
|
Row 10 |
Identifies the ES and ED images on which the analysis is based, for frame by frame analysis the analyzed image are specified in the frame by frame results (3208) template |
The Calibration Template consists of a CONTAINER, with a structure for reporting of the calibration of images used in the analysis.
Extensible
Significant
Table TID 3205. Calibration
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
$CalibrationPlane |
||
|
3 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
MC |
IF different from Analysis program specified in the invoking template |
||
|
4 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
MC |
IF different from Analysis program specified in the invoking template |
||
|
5 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
MC |
IF different from Analysis program specified in the invoking template |
||
|
6 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
7 |
> |
CONTAINS |
CODE |
1 |
MC |
If row 6 value specifies Calibration Object Used |
||
|
8 |
> |
CONTAINS |
NUM |
1 |
MC |
If row 6 value specifies Calibration Object Used |
||
|
9 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (111026, DCM, "Horizontal Pixel Spacing") $Unit = DT (mm/{pixel}, UCUM, "mm/pixel") $ImagePurpose = EV (121112, DCM, "Source of Measurement") |
||
|
10 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (111066, DCM, "Vertical Pixel Spacing") $Unit = DT (mm/{pixel}, UCUM, "mm/pixel") $ImagePurpose = EV (121112, DCM, "Source of Measurement") |
||
|
11 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 3 |
Identifies the Calibration program |
|
Row 4 |
Identifies the Calibration program version |
|
Row 5 |
Identifies the Calibration program manufacturer |
|
Row 7 |
Besides a Sphere and a Catheter, a Distance can be identified as a Calibration Object. In this case a distance measurement of a known dimension of the object is used to calculate the pixel size. |
|
Row 8 |
The catheter size units is also used to specify the size of other calibration objects (e.g., sphere) |
|
Row 9, 10 |
Spacing in the patient body. Point to a single frame containing the image used for calibration if applicable, the actual measurements may be indicated by a SCOORD (see TID 320 “Image or Spatial Coordinates”, row 3) |
|
Row 11 |
Secondary Capture image with calibration position |
The VA Main Results Template consists of a CONTAINER with a structure for reporting the main ventricular analysis measurements.
Extensible
Significant
Table TID 3206. VA Main Results
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
3 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (1, UCUM, "no units") |
||
|
5 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT ({ratio}, UCUM, "ratio") |
||
|
6 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (ml, UCUM, "ml") |
||
|
7 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT ({ratio}, UCUM, "ratio") |
||
|
8 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (ml, UCUM, "ml") |
||
|
9 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = DCID 3467 “Ejection Fraction” $Unit = DT (%, UCUM, "%") |
||
|
10 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3468 “ED Volume” $Unit = DT (ml, UCUM, "ml") |
||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3469 “ES Volume” $Unit = DT (ml, UCUM, "ml") |
||
|
12 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (F-32120, SRT, "Stroke Volume") $Unit = DT (ml, UCUM, "ml") |
||
|
13 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT ({H.B.}/min, UCUM, "BPM") |
||
|
14 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3468 “ED Volume” $ModType = EV (121425, DCM, "Index") $ModValue = DCID 3455 “Index Methods” $Unit = DT (ml/m2, UCUM, "ml/m2") |
||
|
15 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3468 “ED Volume” $ModType = EV (121425, DCM, "Index") $ModValue = EV (29463-7, LN, "Patient Weight") $Unit = DT (ml/kg, UCUM, "ml/kg") |
||
|
16 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3469 “ES Volume” $ModType = EV (121425, DCM, "Index") $ModValue = DCID 3455 “Index Methods” $Unit = DT (ml/m2, UCUM, "ml/m2") |
||
|
17 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3469 “ES Volume” $ModType = EV (121425, DCM, "Index") $ModValue = EV (29463-7, LN, "Patient Weight") $Unit = DT (ml/kg, UCUM, "ml/kg") |
||
|
18 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (F-32120, SRT, "Stroke Volume") $ModType = EV (121425, DCM, "Index") $ModValue = DCID 3455 “Index Methods” $Unit = DT (ml/m2, UCUM, "ml/m2") |
||
|
19 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (F-32120, SRT, "Stroke Volume") $ModType = EV (121425, DCM, "Index") $ModValue = EV (29463-7, LN, "Patient Weight") $Unit = DT (ml/kg, UCUM, "ml/kg") |
||
|
20 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (F-32100, SRT, "Cardiac Output") $Unit = DT (l/min, UCUM, "l/min") |
||
|
21 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (F-32110, SRT, "Cardiac Index") $ModType = EV (121425, DCM, "Index") $ModValue = DCID 3455 “Index Methods” $Unit = DT (l/min/m2, UCUM, "l/min/m2") |
||
|
22 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122445, DCM, "Wall Thickness") $Unit = DT (mm, UCUM, "mm") |
||
|
23 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122446, DCM, "Wall Volume") $Unit = DT (ml, UCUM, "ml") |
||
|
24 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122447, DCM, "Wall Mass") $Unit = DT (g, UCUM, "gram") |
||
|
25 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122447, DCM, "Wall Mass") $ModType = EV (121425, DCM, "Index") $ModValue = DCID 3455 “Index Methods” $Unit = DT (g/m2, UCUM, "gram/m2") |
||
|
26 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122447, DCM, "Wall Mass") $ModType = EV (121425, DCM, "Index") $ModValue = EV (29463-7, LN, "Patient Weight") $Unit = DT (g/kg, UCUM, "gram/kg") |
||
|
27 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122448, DCM, "Wall Stress") $Unit = DT (dyn/cm2, UCUM, "dynes/cm2") |
||
|
28 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1-n |
U |
Content Item Descriptions
|
Row 28 |
Secondary Capture image with ED and/or ES contours |
The AA Main Results Template consists of a CONTAINER with a structure for reporting the main atrial analysis measurements.
Extensible
Significant
Table TID 3207. AA Main Results
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
3 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (1, UCUM, "no units") |
||
|
5 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT ({ratio}, UCUM, "ratio") |
||
|
6 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (ml, UCUM, "ml") |
||
|
7 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT ({ratio}, UCUM, "ratio") |
||
|
8 |
> |
CONTAINS |
NUM |
1 |
U |
Unit = DT (ml, UCUM, "ml") |
||
|
9 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3468 “ED Volume” $Unit = DT (ml, UCUM, "ml") |
||
|
10 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = DCID 3469 “ES Volume” $Unit = DT (ml, UCUM, "ml") |
||
|
11 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1-n |
U |
Content Item Descriptions
|
Row 11 |
Secondary Capture image with ED and/or ES contours |
The Frame-to-Frame Result Template consists of a CONTAINER providing measurements derived from the angiographic images on frame-by-frame basis.
Extensible
Significant
Table TID 3208. Frame-To-Frame Result
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
> |
CONTAINS |
IMAGE |
1-2 |
M |
VM = 1: Single plane analysis, VM = 2: Biplane analysis |
||
|
4 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
5 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = DCID 3471 “Estimated Volumes” $TargetSite = DCID 3462 “Chamber Identification” $Unit = DT (ml, UCUM, "ml") |
||
|
6 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (122445, DCM, "Wall Thickness") $Unit = DT (mm, UCUM, "mm") |
||
|
7 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1-n |
U |
Content Item Descriptions
|
Row 3 |
Identifies each frame analyzed, using the multi-valued Referenced Frame Number (0008,1160) attribute of the IMAGE content item. |
|
Row 5, 6 |
Includes one measurement for each frame referenced in Row 3. |
|
Row 7 |
Secondary Capture image with ventricular contours |
The Centerline Wall Motion Template consists of a CONTAINER providing measurements of the centerline wall motion.
Extensible
Significant
Table TID 3209. Centerline Wall Motion
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
100 |
M |
$Measurement = EV (122450, DCM, "Normalized Chord Length") $Unit = DT (%, UCUM, "%") |
||
|
5 |
> |
CONTAINS |
NUM |
1 |
M |
Values shall be 1, 2 or 3 UNITS = EV ({sd}, UCUM, "Standard Deviations") |
||
|
6 |
> |
CONTAINS |
CONTAINER |
1-6 |
U |
|||
|
7 |
>> |
CONTAINS |
CODE |
EV (F-32050, SRT, "Cardiac Wall Motion") |
1 |
M |
||
|
8 |
>> |
CONTAINS |
CODE |
EV (R-404F0, SRT, "Circumferential Extent") |
1 |
U |
||
|
9 |
>> |
CONTAINS |
NUM |
1 |
M |
Unit = DT (1, UCUM, "no unit") |
||
|
10 |
>> |
CONTAINS |
NUM |
1 |
M |
Unit = DT (1, UCUM, "no unit") |
||
|
11 |
> |
CONTAINS |
CONTAINER |
1-4 |
U |
|||
|
12 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C03E, SRT, "Finding Site") |
1 |
M |
||
|
13 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122459, DCM, "Territory Region Severity") $ModType = EV (F-32050, SRT, "Cardiac Wall Motion") $ModValue = DCID 3703 “Wall Motion” $Unit = DT ({sd}, UCUM, "Standard Deviations") |
||
|
14 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122461, DCM, "Opposite Region Severity") $ModType = EV (F-32050, SRT, "Cardiac Wall Motion") $ModValue = DCID 3703 “Wall Motion” $Unit = DT ({sd}, UCUM, "Standard Deviations") |
||
|
15 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 4 |
Normalized lengths of the chords determined between ED and ES contour. The measurement template allows the specification of the statistical properties of the normal population and of the chord measurement relative to the population. |
|
Row 8 |
If the Circumferential Extent is not specified no limitations to the boundaries for regions are assumed. |
|
Row 11 |
The Regional Abnormal Wall Motion container holds the information on the severity of the decreased or increased wall motion of the 4 predefined regions as described in [Sheehan, 1986]. |
|
Row 12 |
The name of the region with an abnormal ventricular wall motion as described in [Sheehan, 1986]. |
|
Row 13 |
The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the territory region as described in [Sheehan, 1986]. |
|
Row 14 |
The severity of the wall motion abnormality expressed in Standard Deviations above or below normal in the opposite region as described in [Sheehan, 1986]. |
|
Row 15 |
Secondary Capture image with centerline analysis result |
The Radial Based Wall Motion Template consists of a CONTAINER providing measurements of the radial based wall motion.
Extensible
Significant
Table TID 3210. Radial Based Wall Motion
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
CONTAINER |
1-n |
M |
|||
|
5 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
6 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122495, DCM, "Regional Contribution to Ejection Fraction") $Unit = DT (%, UCUM, "%") |
||
|
7 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (122496, DCM, "Radial Shortening") $Unit = DT (%, UCUM, "%") |
||
|
8 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 6 |
The CREF values of the 6 regions determined for the radial based wall motion |
|
Row 7 |
The shortening of the measured radials within the region |
|
Row 8 |
Secondary Capture image with radial based analysis result |
The Landmark Based Wall Motion Template consists of a CONTAINER providing measurements of the landmark based wall motion.
Extensible
Significant
Table TID 3211. Landmark Based Wall Motion
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = EV (122498, DCM, "Slice Contribution to Ejection Fraction") $Unit = DT (%, UCUM, "%") |
||
|
5 |
> |
CONTAINS |
CONTAINER |
5 |
M |
|||
|
6 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Sites") |
1 |
M |
||
|
7 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122495, DCM, "Regional Contribution to Ejection Fraction") $Unit = DT (%, UCUM, "%") |
||
|
8 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 8 |