Quantitative Arterial Analysis Report SR IOD Templates 

The templates that comprise the Quantitative Arterial Analysis SR are interconnected as in Figure A-4:

Quantitative Arterial Analysis Report SR IOD Template Structure

Figure A-4. Quantitative Arterial Analysis Report SR IOD Template Structure


TID 3213 Quantitative Arterial Analysis

The Quantitative Arterial Analysis Template consists of a CONTAINER with a structure for reporting the result of the quantitative arterial analysis process.

Type:

Extensible

Order:

Significant

Table TID 3213. Quantitative Arterial Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122291, DCM, "Quantitative Arteriography Report")

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

4

>

CONTAINS

INCLUDE

DTID 3602 “Cardiovascular Patient Characteristics”

1

U

5

>

HAS OBS CONTEXT

TEXT

EV (111001, DCM, "Algorithm Name")

1

M

6

>

HAS OBS CONTEXT

TEXT

EV (111003, DCM, "Algorithm Version")

1

M

7

>

HAS OBS CONTEXT

TEXT

EV (122405, DCM, "Algorithm Manufacturer")

1

M

8

>

CONTAINS

INCLUDE

DTID 3214 “Analyzed Segment”

1-n

M


Content Item Descriptions

Row 5

Identifies the Arterial Analysis program

Row 6

Identifies the Arterial Analysis program version

Row 7

Identifies the Arterial Analysis program manufacturer

TID 3214 Analyzed Segment

The Analyzed Segment Template consists of a CONTAINER providing quantitative arterial analysis measurements derived from the angiographic images.

Type:

Extensible

Order:

Significant

Table TID 3214. Analyzed Segment

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3604 “Arterial Lesion Locations”

3

>

CONTAINS

IMAGE

EV (121112, DCM, "Source of Measurement")

1

M

4

>

CONTAINS

INCLUDE

DTID 3205 “Calibration”

1

M

5

>

HAS ACQ CONTEXT

INCLUDE

DTID 3520 “Hemodynamic Clinical Context”

1

U

6

>

HAS ACQ CONTEXT

CODE

EV (G-72BB, SRT, "Catheterization Procedure Phase")

1

U

DCID 3651 “Hemodynamic Measurement Phase”

7

>

CONTAINS

SCOORD

EV (122507, DCM, "Left Contour")

1

M

GRAPHIC TYPE = POLYLINE

8

>>

R-SELECTED FROM

IMAGE

1

M

Must reference Row 3

9

>

CONTAINS

SCOORD

EV (122508, DCM, "Right Contour")

1

M

GRAPHIC TYPE = POLYLINE

10

>>

R-SELECTED FROM

IMAGE

1

M

Must reference Row 3

11

>

CONTAINS

INCLUDE

DTID 3219 “Segment Values”

1

M

12

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-404FB, SRT, "Minimum")

$Unit = DT (mm, UCUM, "mm")

13

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (G-A437, SRT, "Maximum")

$Unit = DT (mm, UCUM, "mm")

14

>

CONTAINS

CONTAINER

EV (122509, DCM, "Diameter Graph")

1

U

15

>>

CONTAINS

NUM

EV (122511, DCM, "Graph Increment")

1

M

Value = 1

UNITS = DT ({pixels}, UCUM, "pixels")

16

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Unit = DT (mm, UCUM, "mm")

17

>

CONTAINS

NUM

EV (122382, DCM, "Site of Luminal Minimum ")

1

U

UNITS = DT ({pixels}, UCUM, "pixels")

18

>

CONTAINS

NUM

EV (122516, DCM, "Site of Luminal Maximum")

1

U

UNITS = DT ({pixels}, UCUM, "pixels")

19

>

CONTAINS

INCLUDE

DTID 3215 “Angiographic Lesion Analysis”

1-n

U

20

>

CONTAINS

INCLUDE

DTID 3217 “Sub-segmental Data”

1-n

U

21

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 1

Observation DateTime (0040,A032) of container needs to be flagged with the time of the analysis

Row 6

Numeric coordinates (x,y) identifying the contour points from proximal to distal of left contour. Left is relative to the direction of the blood flow.

Row 9

Numeric coordinates (x,y) identifying the contour points from proximal to distal of right contour. Right is relative to the direction of the blood flow.

Row 12, 13

Positions are relative to the midpoint between the first left and right contour points and measured along the midline between the left and right contour.

Row 14

The X-axis represents the pixel points of the midline of the vessel from proximal to distal. The points on the midline are not necessarily equidistant.

Row 16

For each point of the midline of the vessel a measurement value for the diameter is calculated.

Row 17, 18

The positions in the graph are related to the points on the midline of the vessel.

Row 21

Secondary Capture image with Arterial Analysis contour.

Definition of Left and Right defined by the direction of the blood flow as in Figure A-4b:

Direction of Blood Flow

Figure A-4b. Direction of Blood Flow


TID 3215 Angiographic Lesion Analysis

The Angiographic Lesion Analysis Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.

Type:

Extensible

Order:

Significant

Table TID 3215. Angiographic Lesion Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (F-00585, SRT, "Lesion Finding")

1

M

2

>

CONTAINS

TEXT

EV (121151, DCM, "Lesion Identifier")

1

M

3

>>

HAS PROPERTIES

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3604 “Arterial Lesion Locations”

4

>>>

HAS CONCEPT MOD

CODE

EV (G-A1F8, SRT, "Topographical modifier")

1

U

DCID 3019 “Cardiovascular Anatomic Location Modifiers”

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-404FB, SRT, "Minimum")

$Unit = DT (mm, UCUM, "mm")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area")

$Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods”

$Derivation = EV (R-404FB, SRT, "Minimum")

$Unit = DT (mm2, UCUM, "mm2")

7

>

CONTAINS

CODE

EV (122430, DCM, "Reference Method")

1

M

DCID 3465 “QA Reference Methods”

8

>

CONTAINS

CONTAINER

EV (122438, DCM, "Reference Points")

1

U

9

>>

CONTAINS

NUM

EV (122337, DCM, "Relative Position")

1-n

M

UNITS = DT (mm, UCUM, "mm")

10

>>>

HAS PROPERTIES

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Unit = DT (mm, UCUM, "mm")

11

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$TargetSite = (122382, DCM, "Site of Luminal Minimum")

$Unit = DT (mm, UCUM, "mm")

12

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area")

$Derivation = EV (122404, DCM, "Reconstructed")

$TargetSite = (122382, DCM, "Site of Luminal Minimum")

$Unit = DT (mm2, UCUM, "mm2")

13

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-41D2D, SRT, "Calculated")

$TargetSite = EV (122481, DCM, "Contour Start")

$Unit = DT (mm, UCUM, "mm")

14

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-41D2D, SRT, "Calculated")

$TargetSite = EV (122482, DCM, "Contour End")

$Unit = DT (mm, UCUM, "mm")

15

>

CONTAINS

INCLUDE

DTID 3218 “Position in Arterial Segment”

1

M

16

>

CONTAINS

CONTAINER

EV (122517, DCM, "Densitometric Luminal Cross-sectional Area Graph")

1

U

17

>>

CONTAINS

NUM

EV (122511, DCM, "Graph Increment")

1

M

Value = 1

UNITS = DT ({pixels}, UCUM, "pixels")

18

>>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area")

$Unit = (mm2, UCUM, "mm2")

19

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area")

$Derivation = EV (R-41D2D, SRT, "Calculated")

$Method = EV (122474, DCM, "Densitometric method")

$TargetSite = EV (122481, DCM, "Contour Start")

$Unit = (mm2, UCUM, "mm2")

20

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area")

$Derivation = EV (R-41D2D, SRT, "Calculated")

$Method = EV (122474, DCM, "Densitometric method")

$TargetSite = EV (122482, DCM, "Contour End")

$Unit = (mm2, UCUM, "mm2")

21

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (R-101BC, SRT, "Stenotic Lesion Length")

$Unit = DT (mm, UCUM, "mm")

22

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (R-101BB, SRT, "Lumen Diameter Stenosis")

$Unit = DT (%, UCUM, "%")

23

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (R-101BA, SRT, "Lumen Area Stenosis")

$Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods”

$Unit = DT (%, UCUM, "%")

24

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (122372, DCM, "Lumen Volume")

$Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods”

$Unit = DT (mm3, UCUM, "mm3")

25

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122542, DCM, "Plaque Area")

$Unit = DT (mm2, UCUM, "mm2")

26

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122376, DCM, "Total Plaque Volume")

$Unit = DT (mm3, UCUM, "mm3")

27

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122544, DCM, "Diameter Symmetry")

$Unit = DT ({ratio}, UCUM, "ratio")

28

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122545, DCM, "Area Symmetry")

$Unit = DT ({ratio}, UCUM, "ratio")

29

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122546, DCM, "Inflow Angle")

$Unit = DT (deg, UCUM, "deg")

30

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122547, DCM, "Outflow Angle")

$Unit = DT (deg, UCUM, "deg")

31

>

CONTAINS

INCLUDE

DTID 3216 “Stenotic Flow Reserve”

1

U

32

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 8

Set of user defined reference position for method that requires local reference position.

Row 9

Distance of local reference position from an arbitrary landmark.

Row 10

Diameter at a local reference position.

Row 11

The reference diameter for the arterial lesion calculated with the applicable reference method

Row 12

The reference area for the arterial lesion calculated with the applicable reference method

Row 13

The diameter measurement at the start of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14)

Row 14

The diameter measurement at the end of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14)

Row 15

The positions of the lesion, borders of the lesion, etc.

Row 16

The graph with the calculated cross sectional area results based on the densitometric method

Row 18

The cross sectional area measurements calculated based on the densitometric method

Row 19

The cross sectional area measurement at the start of the reconstruction line in the area graph

Row 20

The cross sectional area measurement at the end of the reconstruction line in the area graph

Row 21

Measured along the midline of the left and right contour

Row 22

The diameter stenosis is calculated as follows:

(Reference Luminal Diameter - Minimum Luminal Diameter / Reference Luminal Diameter) * 100%

Row 23

The circular and the densitometric area stenosis are calculated respectively as:

(Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Circular Area / Reference Vessel Lumen Cross-Sectional Area) * 100%(Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Densitometric Area / Reference Vessel Lumen Cross-Sectional Area) * 100%

Row 24

Estimated lumen volume between proximal border and distal border of lesion (TID 3218 “Position in Arterial Segment”, row 1 and 2)

Row 32

Secondary Capture image with obstruction analysis contour

TID 3216 Stenotic Flow Reserve

The Stenotic Flow Reserve Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.

Type:

Extensible

Order:

Significant

Table TID 3216. Stenotic Flow Reserve

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122548, DCM, "Stenotic Flow Reserve")

$Unit = DT ({ratio}, UCUM, "ratio")

2

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122549, DCM, "Poiseuille Resistance")

$Unit = DT (mm[Hg].s/cm, UCUM, "mmHG.s/cm")

3

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122550, DCM, "Turbulence Resistance")

$Unit = DT (mm[Hg].s2/cm2, UCUM, "mmHG.s2/cm2")

4

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122555, DCM, "Estimated Normal Flow")

$Unit = DT (ml/s, UCUM, "ml/s")

5

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122551, DCM, "Pressure Drop at SFR")

$Unit = DT (mm[Hg], UCUM, "mmHg")

6

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 6

Secondary Capture image with SFR analysis contour

TID 3217 Sub-segmental Data

The Sub-segmental Data Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for a sub-segment in a total analyzed segment.

Type:

Extensible

Order:

Significant

Table TID 3217. Sub-Segmental Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (121070, DCM, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (G-C0E3, SRT, "Finding Site")

1

M

DCID 3604 “Arterial Lesion Locations”

3

>>

HAS CONCEPT MOD

CODE

EV (G-A1F8, SRT, "Topographical modifier")

1

U

DCID 3019 “Cardiovascular Anatomic Location Modifiers”

4

>

CONTAINS

CODE

EV (122554, DCM, "Segmentation Method")

1

M

DCID 3456 “Sub-segment Methods”

5

>

CONTAINS

INCLUDE

DTID 3219 “Segment Values”

1

U

6

>

CONTAINS

INCLUDE

DTID 3218 “Position in Arterial Segment”

1

M

7

>

CONTAINS

IMAGE

No purpose of reference

1

U


Content Item Descriptions

Row 7

Secondary Capture image with obstruction analysis contour

TID 3218 Position in Arterial Segment

The Position in Arterial Segment Template consists of the position content items common for the Angiographic Lesion Analysis and Sub-Segmental Data.

Type:

Extensible

Order:

Significant

Table TID 3218. Position in Arterial Segment

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122528, DCM, "Position of Proximal Border")

$Unit = DT (mm, UCUM, "mm")

2

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122529, DCM, "Position of Distal Border")

$Unit = DT (mm, UCUM, "mm")

3

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122382, DCM, "Site of Luminal Minimum")

$Unit = DT (mm, UCUM, "mm")

4

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122516, DCM, "Site of Luminal Maximum")

$Unit = DT (mm, UCUM, "mm")

5

CONTAINS

NUM

EV (122528, DCM, "Position of Proximal Border")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

6

CONTAINS

NUM

EV (122529, DCM, "Position of Distal Border")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

7

CONTAINS

NUM

EV (122382, DCM, "Site of Luminal Minimum")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")

8

CONTAINS

NUM

EV (122516, DCM, "Site of Luminal Maximum")

1

UC

IFF TID 3214 “Analyzed Segment” Row 14 is present

UNITS = DT ({pixels}, UCUM, "pixels")


Content Item Descriptions

Row 1, 2, 3, 4

Positions are relative to the midpoint of the first left and right contour points and measured along the midline of the left and right contour

Row 5, 6, 7, 8

The positions are relative to the measurement locations of the Diameter Graph of TID 3214 “Analyzed Segment” row 14.

TID 3219 Segment Values

The Segment Values Template consists of content items providing quantitative arterial analysis measurements for a total analyzed segment or sub segment.

Type:

Extensible

Order:

Significant

Table TID 3219. Segment Values

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122510, DCM, "Length Luminal Segment")

$Unit = DT (mm, UCUM, "mm")

2

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-404FB, SRT, "Minimum")

$Unit = DT (mm, UCUM, "mm")

3

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (G-A437, SRT, "Maximum")

$Unit = DT (mm, UCUM, "mm")

4

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-00317, SRT, "Mean")

$Unit = DT (mm, UCUM, "mm")

5

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter")

$Derivation = EV (R-10047, SRT, "Standard Deviation")

$Unit = DT (mm, UCUM, "mm")


Content Item Descriptions

Row 1

Measured along the midline of the left and right contour.