The ECG Report template is the root structure for the representation of analysis of an ECG waveform, potentially in comparison to a prior ECG waveform analysis. The analyzed waveform may or may not be stored as a DICOM SOP Instance.
Extensible
Significant
Yes
Table TID 3700. ECG Report
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
Root node |
||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
|||
|
3 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
|||
|
5 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
|
6 |
>> |
CONTAINS |
CODE |
1-n |
U |
|||
|
7 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
|
8 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
9 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
10 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
12 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
|
13 |
>> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
14 |
>> |
CONTAINS |
INCLUDE |
1-n |
U |
One instantiation per reported lead |
||
|
15 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
16 |
> |
CONTAINS |
INCLUDE |
1 |
U |
Extensible
Significant
Table TID 3702. Prior ECG Exam
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
|
2 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
3 |
> |
CONTAINS |
DATETIME |
1 |
U |
|||
|
4 |
> |
CONTAINS |
UIDREF |
1 |
U |
|||
|
5 |
> |
CONTAINS |
COMPOSITE |
1 |
U |
|||
|
6 |
> |
CONTAINS |
WAVEFORM |
1 |
U |
Extensible
Significant
Table TID 3704. Patient Characteristics for ECG
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = DCID 7456 “Units of Measure for Age” |
||
|
3 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (cm, UCUM, "cm") |
||
|
5 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (kg, UCUM, "kg") |
||
|
6 |
> |
CONTAINS |
NUM |
EV (F-008EC, SRT, "Systolic Blood Pressure") |
1 |
U |
UNITS = DCID 3500 “Pressure Units” |
|
|
7 |
> |
CONTAINS |
NUM |
EV (F-008ED, SRT, "Diastolic Blood Pressure") |
1 |
U |
UNITS = DCID 3500 “Pressure Units” |
|
|
8 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
9 |
> |
CONTAINS |
CODE |
EV (R-00728, SRT, "Pacemaker in situ") |
1 |
U |
||
|
10 |
> |
CONTAINS |
CODE |
EV (R-0077C, SRT, "ICD in situ") |
1 |
U |
The ECG Waveform Information template provides reference to, and important parameters of, the analyzed waveform.
Extensible
Significant
Table TID 3708. ECG Waveform Information
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
CONTAINS |
WAVEFORM |
1 |
U |
|||
|
3 |
> |
CONTAINS |
CODE |
EV (10:11345, MDC, "Lead System") |
1 |
U |
||
|
4 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
5 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
6 |
> |
CONTAINS |
INCLUDE |
1 |
U |
Person performing the ECG acquisition |
||
|
7 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
8 |
> |
CONTAINS |
DATETIME |
1 |
M |
|||
|
9 |
> |
CONTAINS |
NUM |
1-n |
U |
|||
|
10 |
> |
CONTAINS |
TEXT |
1-n |
U |
|||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
U |
The ECG Global Measurements Template provides a structure for measurements calculated across the ECG waveform as a whole (multiple leads).
As an Extensible Template, applications may include any ECG global measurements, such as angles of the electrical vector of various ECG waves. The recommended vocabulary for such concepts is ISO/IEEE 11073-10102.
Extensible
Significant
Table TID 3713. ECG Global Measurements
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
|
3 |
> |
CONTAINS |
NUM |
DT (2:16020, MDC, "Atrial Heart Rate") |
1 |
U |
UNITS = EV ({H.B.}/min, UCUM, "BPM") |
|
|
4 |
> |
CONTAINS |
NUM |
DT (2:16016, MDC, "Ventricular Heart Rate") |
1 |
M |
UNITS = EV ({H.B.}/min, UCUM, "BPM") |
|
|
5 |
> |
CONTAINS |
NUM |
DT (2:16160, MDC, "QT interval global") |
1 |
M |
UNITS = EV (ms, UCUM, "ms") |
|
|
6 |
> |
CONTAINS |
NUM |
DT (2:15876, MDC, "QTc interval global") |
1 |
U |
UNITS = EV (ms, UCUM, "ms") |
|
|
7 |
>> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
|
8 |
> |
CONTAINS |
NUM |
DT (2:15872, MDC, "PR interval global") |
1 |
M |
UNITS = EV (ms, UCUM, "ms") |
|
|
9 |
> |
CONTAINS |
NUM |
DT (2:16156, MDC, "QRS duration global") |
1 |
M |
UNITS = EV (ms, UCUM, "ms") |
|
|
10 |
> |
CONTAINS |
NUM |
DT (2:16168, MDC, "RR interval global") |
1 |
M |
UNITS = EV (ms, UCUM, "ms") |
|
|
11 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV (ms, UCUM, "ms") |
||
|
12 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV (deg, UCUM, "deg") |
||
|
13 |
> |
CONTAINS |
NUM |
DT (2:16032, MDC, "Count of all beats") |
1 |
U |
UNITS = EV ({beats}, UCUM, "beats") |
|
|
14 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV ({beats}, UCUM, "beats") |
||
|
15 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C504, SRT, "Associated Morphology") |
1-n |
U |
The ECG Lead Measurements Template provides a structure for measurements calculated on individual ECG leads.
As an Extensible Template, applications may include any ECG per lead measurements, such as integrals over time of various ECG wave voltages. The recommended vocabulary for such concepts is ISO/IEEE 11073-10102.
Extensible
Significant
Table TID 3714. ECG Lead Measurements
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
|
4 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV (ms, UCUM, "ms") |
||
|
5 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV (mV, UCUM, "mV") |
||
|
6 |
> |
CONTAINS |
CODE |
EV (F-38035, SRT, "ST Segment Finding") |
1 |
U |
||
|
7 |
> |
CONTAINS |
CODE |
1-n |
U |
The ECG Measurement Source Template provides a structure for identifying the particular cardiac cycle, or beat, in an analyzed ECG waveform used for the measurement group for which this template provides Observation Context. The cardiac cycle is identified by beat number, and optionally by specific temporal coordinates within a DICOM ECG Waveform SOP Instance.
Extensible
Significant
Table TID 3715. ECG Measurement Source
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
TEXT |
1 |
U |
Up to three numeric characters |
||||
|
2 |
CODE |
EV (G-C036, SRT, "Measurement Method") |
1 |
U |
||||
|
3 |
TCOORD |
1 |
U |
|||||
|
4 |
> |
SELECTED FROM |
WAVEFORM |
1 |
U |
Content Item Descriptions
|
Row 1 |
Beat Number is specified as a numeric text string, and shall be treated as the ordinal of the beat (cardiac cycle) within the waveform acquisition for this lead that was analyzed for the measurements in this container (i.e., "1" for the first beat, "2" for the second, etc.). If absent, the measurements may have been made by a technique across multiple cycles as specified in Row 2 Measurement Method. |
|
Rows 3 and 4 |
Source of measurement identify the specific channel and time period within a DICOM ECG Waveform SOP Instance that was analyzed for the measurements in this container. |
The ECG Qualitative Analysis template allows a free text qualitative interpretation of the analyzed ECG, as well as a structure for a coded analysis.
Extensible
Significant
Table TID 3717. Qualitative Analysis, ECG
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINS |
CONTAINER |
1 |
M |
||||
|
2 |
> |
CONTAINS |
TEXT |
EV (F-00033, SRT, "ECG Finding") |
1 |
MC |
At least one of rows 2 and 3 shall be present |
|
|
3 |
> |
CONTAINS |
CODE |
EV (F-00033, SRT, "ECG Finding") |
1-n |
MC |
At least one of rows 2 and 3 shall be present |
|
|
4 |
>> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
Interpretive statement |
||
|
5 |
>> |
INFERRED FROM |
CODE |
EV (F-00033, SRT, "ECG Finding") |
1-n |
U |
No BCID; may use implementation-specific codes |
|
|
6 |
>>> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
Interpretive statement |
Content Item Descriptions
|
Row 3-4 |
ECG Finding provides one or more coded interpretive statements using standard or implementation-specific codes. Each coded finding will include a Code Meaning (0008,0104) using the LO Value Representation (64 characters); longer human-readable text strings may be conveyed in the Row 4 Equivalent Meaning of Value Content Item.. |
|
Row 5-6 |
Each primary ECG Finding of Row 3 may have multiple supporting coded findings in Row 5, with longer human-readable text strings if necessary in Row 6. |