The templates that comprise the Mammography CAD SR IOD are interconnected as in Figure A-8:
In Figure A-8, '···' indicates possible recursive application of subordinate templates.
This template forms the top of a content tree that allows a mammography CAD device to describe the results of detection and analysis of Mammographic evidence. This template, together with its subordinate templates, describes both the results for presentation to radiologists and partial product results for consumption by mammography CAD devices in subsequent mammography CAD reports.
This template defines a Container that contains an Image Library, the mammography CAD results, and summaries of the detection and analysis algorithms performed. The Image Library contains the Image SOP Class and Instance UIDs, and selected attributes for each image referenced in either the algorithm summaries or mammography CAD results.
The Summary of Detections and Summary of Analyses sub-trees gather lists of algorithms attempted, grouped by success/failure status. Algorithms not attempted are not mentioned in these sub-trees. This information forms the basis for understanding why a mammography CAD report may produce no (or fewer than anticipated) results. Mammography CAD results are constructed bottom-up, starting from Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”), associated as Composite Features (see TID 4004 “Mammography CAD Composite Feature”), and from which Individual and Overall Impressions are formed.
See Figure E.1-1 “Top Levels of Mammography CAD SR Content Tree” in PS3.17 .
Non-Extensible
Significant
Yes
Table TID 4000. Mammography CAD Document Root
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
Root node |
||||
|
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
|
3 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
|
4 |
>> |
CONTAINS |
INCLUDE |
1-n |
M |
$ImageLaterality = DCID 6022 “Side” $ImageView = DCID 4014 “View for Mammography” $ImageViewMod = DCID 4015 “View Modifier for Mammography” |
||
|
5 |
> |
CONTAINS |
INCLUDE |
DTID 4001 “Mammography CAD Overall Impression/Recommendation” |
1 |
M |
||
|
6 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
7 |
>> |
INFERRED FROM |
INCLUDE |
1 |
MC |
Shall be present unless the value of (111064, DCM, "Summary of Detections") is (111225, DCM, "Not Attempted") |
$DetectionCode = DCID 6014 “Mammography Single Image Finding” |
|
|
8 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
9 |
>> |
INFERRED FROM |
INCLUDE |
1 |
MC |
Shall be present unless the value of (111065, DCM, "Summary of Analyses") is (111225, DCM, "Not Attempted") |
$AnalysisCode = DCID 6043 “Types of Mammography CAD Analysis” |
Content Item Descriptions
|
Image Library |
The "Image Library" section of the Content Tree (TID 4000 “Mammography CAD Document Root”, row 3) shall include all Image SOP Instances from the Current Requested Procedure Evidence Sequence (0040,A375) attribute of the SR Document General module. If a portion of another instance of a Mammography CAD SR IOD is duplicated in the "Overall Impression/ Recommendation" section of the Content Tree, the "Image Library" shall also include all Image Library Entries referenced from the duplicated portions of the Mammography CAD SR. |
|
Detections Performed |
The "Detections Performed" and "Analyses Performed" sections of the Content Tree (TID 4000 “Mammography CAD Document Root”, rows 6 and 8) together shall reference all Image SOP Instances included in the Current Requested Procedure Evidence Sequence (0040,A375) attribute of the SR Document General module. |
|
Analyses Performed |
This template forms the top of the mammography CAD results sub-tree. The contents of this template describe the overall impression the mammography CAD device had for the mammographic evidence presented and any recommendations that the mammography CAD device made. The details of the overall impression and recommendation are expressed in this instance of the Mammography CAD Impression/Recommendation Body (see TID 4002 “Mammography CAD Impression/Recommendation Body”). The data from which the details are inferred, are expressed in the Mammography CAD Individual Impression/Recommendations (see TID 4003 “Mammography CAD Individual Impression/Recommendation”), of which there may be several.
Non-Extensible
Significant
Table TID 4001. Mammography CAD Overall Impression/Recommendation
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
|||||
|
2 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
|||
|
3 |
> |
INFERRED FROM |
INCLUDE |
DTID 4003 “Mammography CAD Individual Impression/Recommendation” |
1-n |
MC |
Shall be present if 1 or more (111059, DCM, "Single Image Finding") or (111015, DCM, "Composite Feature") content items are reported. |
Content Item Descriptions
|
CAD Processing and Findings Summary |
This code value is used to express if and why the Overall Impression/Recommendation sub-tree is empty. The Summary of Detections and Summary of Analyses sub-trees of the Document Root node contain detail about which (if any) algorithms succeeded or failed. If the code value indicates that there were no findings, then the code value can be used to determine whether mammography CAD processing occurred successfully, without parsing the Summary of Detections and Summary of Analyses sub-trees. |
|
Row 3 |
There are no constraints regarding the 1-n multiplicity of the inclusion of TID 4003 “Mammography CAD Individual Impression/Recommendation” or its underlying structure, other than the TID 4001 “Mammography CAD Overall Impression/Recommendation” and TID 4003 “Mammography CAD Individual Impression/Recommendation” requirements. Individual Impression/Recommendation containers may be organized, for example per image, per finding or composite feature, or some combination thereof. |
The details of an impression and recommendation are expressed in this template. It is applied to both Mammography CAD Overall Impression/Recommendation (TID 4001 “Mammography CAD Overall Impression/Recommendation”) and Mammography CAD Individual Impression/Recommendation (TID 4003 “Mammography CAD Individual Impression/Recommendation”).
Non-Extensible
Significant
Table TID 4002. Mammography CAD Impression/Recommendation Body
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1-n |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. |
||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
|
3 |
CODE |
1-n |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. |
||||
|
4 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
|
5 |
TEXT |
1 |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. |
||||
|
6 |
CODE |
1-n |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. |
||||
|
7 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
|
8 |
NUM |
1 |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. May be present only if (111054, DCM, "Recommended Follow-up Date") is not present. |
UNITS = DCID 6046 “Units of Follow-up Interval” Values = Integer ≥ 0, where 0 = immediate follow-up |
|||
|
9 |
DATE |
1 |
MC |
At least one of rows 1, 3, 5, 6, 8, 9 shall be present. May be present only if (111055, DCM, "Recommended Follow-up Interval") is not present. |
Shall be later than date of exam |
|||
|
10 |
NUM |
1 |
UC |
May be present only if (111005, DCM, "Assessment Category"), (111023, DCM, "Differential Diagnosis/Impression") or (111033, DCM, "Impression Description") is present. |
UNITS = EV (%, UCUM, "Percent") Values = 0 - 100 |
|||
|
11 |
INCLUDE |
1-n |
M |
|||||
|
12 |
NUM |
1-n |
U |
|||||
|
13 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
|
14 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
15 |
> |
INFERRED FROM |
TEXT |
1 |
U |
Content Item Descriptions
|
Certainty of Impression |
The certainty that the device populating the Mammography CAD SR report places on this impression, where 0 equals no certainty and 100 equals certainty. |
|
Impression Description |
Free-form text describing the overall or an individual impression |
This template collects an individual impression the mammography CAD device had for a lesion, non-lesion object, or correlation of related objects. The details of the impression and recommendation are expressed in the Mammography CAD Impression/Recommendation Body (see TID 4002 “Mammography CAD Impression/Recommendation Body”). The data from which the details are inferred are expressed in the Composite Features (see TID 4004 “Mammography CAD Composite Feature”) and/or Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”) of which there may be several.
The sub-tree headed by this template is illustrated in Figure E.1-3 “Example of Individual Impression/Recommendation Levels of Mammography CAD SR Content Tree” in PS3.17 .
Non-Extensible
Significant
Table TID 4003. Mammography CAD Individual Impression/Recommendation
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
MC |
At least one of rows 4, 5 shall be present. |
||
|
5 |
> |
CONTAINS |
INCLUDE |
1-n |
MC |
At least one of rows 4, 5 shall be present. |
Content Item Descriptions
|
Rendering Intent |
This content item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this template and its target content items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |
This template collects a composite feature for a lesion, non-lesion object, or correlation of related objects. The details of the composition are expressed in the Mammography CAD Composite Feature Body (see TID 4005 “Mammography CAD Composite Feature Body”). The data from which the details are inferred, are expressed in the Composite Features (see TID 4004 “Mammography CAD Composite Feature”) and/or Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”), of which there may be several.
A Composite Feature shall be INFERRED FROM any combination of two or more Composite Features or Single Image Findings or mixture thereof.
Non-Extensible
Significant
Table TID 4004. Mammography CAD Composite Feature
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
|
4 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
M |
|||
|
5 |
> |
INFERRED FROM |
INCLUDE |
1-n |
MC |
At least two items shall be present: two of row 5, two of row 6, or one of each. |
||
|
6 |
> |
INFERRED FROM |
INCLUDE |
1-n |
MC |
At least two items shall be present: two of row 5, two of row 6, or one of each. |
||
|
7 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
Shall be present only if this feature is incorporated from a different report than its parent. |
Content Item Descriptions
|
Rendering Intent |
This content item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this template and its target content items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |
The details of a composite feature are expressed in this template. It is applied to Mammography CAD Composite Feature (TID 4004 “Mammography CAD Composite Feature”).
Non-Extensible
Significant
Table TID 4005. Mammography CAD Composite Feature Body
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
DCID 6035 “Composite Feature Relations” The value shall be (111155, DCM, "Target content items are related contra-laterally") if the parent content item has code value (F-01792, SRT, "Focal asymmetric breast tissue") or (F-01793, SRT, "Asymmetric breast tissue"). |
||||
|
2 |
CODE |
1 |
M |
|||||
|
3 |
INCLUDE |
1 |
M |
|||||
|
4 |
NUM |
1 |
U |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
||||
|
5 |
NUM |
1 |
UC |
May be present only if value of parent is not (111102, DCM, "Non-lesion") |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
|||
|
6 |
CODE |
1-n |
U |
|||||
|
7 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) "Linear Measurement" shall be used. |
||||
|
8 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) "Area Measurement" shall be used. |
||||
|
9 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) "Volume Measurement" shall be used. |
||||
|
10 |
INCLUDE |
1-n |
U |
|||||
|
11 |
NUM |
DCID 6037 “Mammography Quantitative Temporal Difference Type” |
1-n |
UC |
May be present only if the value of (111016, DCM, "Composite type") is (111153, DCM, "Target content items are related temporally") |
UNITS = DCID 7460 “Units of Linear Measurement” UNITS = DCID 7461 “Units of Area Measurement” UNITS = DCID 7462 “Units of Volume Measurement” UNITS = DT (1, UCUM, "no units") |
||
|
12 |
> |
R-INFERRED FROM |
NUM |
2 |
U |
The referenced numeric values shall have the same Concept Name. Their UNITS shall be the same as row 11 |
||
|
13 |
CODE |
1-n |
UC |
May be present only if the value of (111016, DCM, "Composite type") is (111153, DCM, "Target content items are related temporally") |
DCID 6038 “Mammography Qualitative Temporal Difference Type” |
|||
|
14 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
|
15 |
> |
R-INFERRED FROM |
CODE |
2 |
M |
The referenced code values shall have the same Concept Name and be from the same context group. |
||
|
16 |
CODE |
1 |
U |
|||||
|
17 |
CODE |
1 |
U |
|||||
|
18 |
CODE |
1 |
U |
|||||
|
19 |
CODE |
1 |
UC |
May be present only if value of parent is (F-01791, SRT, "Mammographic breast mass") or (F-01796, SRT, "Mammography breast density") |
||||
|
20 |
CODE |
EV (M-020F9, SRT, "Shape") |
1 |
UC |
May be present only if value of parent is (F-01791, SRT, "Mammographic breast mass") or (F-01796, SRT, "Mammography breast density") |
|||
|
21 |
CODE |
1-n |
UC |
May be present only if value of parent is (F-01791, SRT, "Mammographic breast mass") or (F-01796, SRT, "Mammography breast density") |
||||
|
22 |
CODE |
1-n |
UC |
May be present only if value of parent is (F-01775, SRT, "Calcification Cluster") or (F-01776, SRT, "Individual Calcification") |
||||
|
23 |
CODE |
1 |
UC |
May be present only if value of parent is (F-01775, SRT, "Calcification Cluster") |
||||
|
24 |
NUM |
1 |
UC |
May be present only if value of parent is (F-01775, SRT, "Calcification Cluster") |
UNITS = EV (1, UCUM, "no units") Value = Integer 1 - n |
|||
|
25 |
NUM |
1-n |
U |
|||||
|
26 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
27 |
> |
INFERRED FROM |
TEXT |
1 |
U |
Content Item Descriptions
|
Certainty of Feature |
The likelihood that the feature analyzed, and classified by the CODE specified in the Composite Feature parent template, is in fact that type of feature. |
|
Volume Measurement |
If dimensions for a volume are to be stated in terms of length, width, and depth, then one shall use 3 instances of TID (1400) Linear Measurement. |
|
Row 11 |
Values ≤ 0 are allowed. The two referenced numeric values are target content items of the first generation Composite Feature or Single Image Finding children of this composite feature. Given the equation, A - B, the value representing A shall be referenced first. |
|
Qualitative Difference |
The two referenced code values are target content items of the first generation Composite Feature or Single Image Finding children of this composite feature. |
This template describes a single image finding for a lesion or other object. The details of the finding are expressed in this template and/or more specific templates. The details from which a single image Calcification Cluster is inferred may be expressed in a number of Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”) of type Individual Calcification.
A Single Image Finding of type Breast Composition may be INFERRED FROM by-reference to a Single Image Finding of type Breast Geometry.
Non-Extensible
Significant
Table TID 4006. Mammography CAD Single Image Finding
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
>> |
HAS PROPERTIES |
NUM |
1 |
UC |
IFF value of row 2 is (111151, DCM, "Presentation Optional") and row 9 of TID 4017 “CAD Detection Performed” is present |
UNITS = DT ({1:n}, UCUM, "range: 1:n"), where n is the maximum specified in Row 9 of TID 4017 “CAD Detection Performed”. Value is restricted to being an integer |
|
|
4 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
|
5 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
M |
|||
|
6 |
> |
HAS PROPERTIES |
NUM |
1 |
U |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
||
|
7 |
> |
HAS PROPERTIES |
NUM |
1 |
UC |
May be present unless value of parent is (F-01710, SRT, "Breast composition"), (111100, DCM, "Breast geometry"), (T-04100, SRT, "Nipple"), (111099, DCM, "Selected region"), (111101, DCM, "Image quality") or (111102, DCM, "Non-lesion") |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
|
|
8 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
Shall be present unless value of parent is (F-01710, SRT, "Breast composition"), (111100, DCM, "Breast geometry") or (111101, DCM, "Image quality") |
||
|
9 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
Shall be present only if value of parent is (F-01710, SRT, "Breast composition") |
||
|
10 |
> |
R-INFERRED FROM |
CODE |
1-n |
UC |
May be present only if value of parent is (F-01710, SRT, "Breast composition") |
Shall reference a (111059, DCM, "Single Image Finding") of value: EV (111100, DCM, "Breast geometry") |
|
|
11 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
Shall be present only if value of parent is (111100, DCM, "Breast geometry") |
||
|
12 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
UC |
May be present only if value of parent is (F-01776, SRT, "Individual Calcification") |
||
|
13 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
UC |
May be present only if value of parent is (F-01775, SRT, "Calcification Cluster") |
||
|
14 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
UC |
May be present only if value of parent is (F-01796, SRT, "Mammography breast density") |
||
|
15 |
> |
HAS PROPERTIES |
CODE |
1 |
UC |
May be present only if value of parent is (T-04100, SRT, "Nipple") |
||
|
16 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
Shall be present only if value of parent is (111102, DCM, "Non-lesion") |
||
|
17 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
Shall be present only if value of parent is (111099, DCM, "Selected Region") |
||
|
18 |
> |
R-INFERRED FROM |
IMAGE |
1 |
MC |
IF value of parent is (111101, DCM, "Image quality") and IFF row 19 is not present |
Shall reference an IMAGE content item in the (111028, DCM, "Image Library") |
|
|
19 |
> |
HAS PROPERTIES |
SCOORD |
1-n |
MC |
IF value of parent is (111101, DCM, "Image quality") and IFF row 18 is not present |
||
|
20 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
M |
All the (111030, DCM, "Image Region") content items in a single invocation of this template shall reference the same IMAGE content item in the (111028, DCM, "Image Library") |
||
|
21 |
> |
HAS PROPERTIES |
INCLUDE |
1-n |
MC |
Shall be present only if value of parent is (111101, DCM, "Image quality") |
$QualityFinding = DCID 6041 “Mammography Image Quality Finding” $QualityStandard = DCID 6045 “Mammography Types of Quality Control Standard” |
|
|
22 |
> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
|
23 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
24 |
>> |
INFERRED FROM |
TEXT |
1 |
U |
|||
|
25 |
> |
INFERRED FROM |
INCLUDE |
1-n |
UC |
May be present only if value of parent is (F-01775, SRT, "Calcification Cluster") |
EV (F-01776, SRT, "Individual Calcification") |
|
|
26 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
Shall be present only if this finding is incorporated from a different report than its parent. |
Content Item Descriptions
|
Rendering Intent |
This content item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this template and its target content items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |
|
CAD Operating Point |
Additional information to use when Rendering Intent is "Presentation Optional". A CAD Operating Point of zero is not sent, and is encoded as a Rendering Intent of "Presentation Required". See Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 and Section E.4 “CAD Operating Point” in PS3.17 . |
|
Single Image Finding |
A Single Image Finding (whose parent is a Single Image Finding of type Calcification Cluster) allows one level of nesting for the definition of individual calcifications within the cluster. To use this template recursively, this Single Image Finding code value shall be "Individual Calcification". |
|
Certainty of Finding |
The likelihood that the finding detected, and classified by the CODE specified in the Single Image Finding parent template, is in fact that type of finding. |
Non-Extensible
Significant
Table TID 4007. Mammography CAD Breast Composition
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
EV (F-01710, SRT, "Breast composition") |
1 |
MC |
At least one of row 1 or 2 shall be present |
|||
|
2 |
NUM |
1 |
MC |
At least one of row 1 or 2 shall be present |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
Content Item Descriptions
|
Percent Fibroglandular Tissue |
Percent of breast area that is mammographically dense, excluding pectoralis muscle. |
Non-Extensible
Significant
Table TID 4008. Mammography CAD Breast Geometry
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
SCOORD |
EV (111007, DCM, "Breast Outline Including Pectoral Muscle Tissue") |
1 |
M |
GRAPHIC TYPE = {POLYLINE} |
|||
|
2 |
> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Shall reference an IMAGE content item in the (111028, DCM, "Image Library") |
||
|
3 |
SCOORD |
1 |
U |
GRAPHIC TYPE = {POLYLINE} |
||||
|
4 |
> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Shall reference the same node as row 2 |
Content Item Descriptions
This template provides the detail specific to an individual calcification.
Non-Extensible
Significant
Table TID 4009. Mammography CAD Individual Calcification
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1-n |
MC |
At least one of rows 1, 2, 3 shall be present |
||||
|
2 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) Linear Measurement shall be used. |
|||
|
3 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) Area Measurement shall be used. |
|||
|
4 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) Volume Measurement shall be used. |
This template provides the detail specific to a calcification cluster.
Non-Extensible
Significant
Table TID 4010. Mammography CAD Calcification Cluster
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
||||
|
2 |
CODE |
1 |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
||||
|
3 |
NUM |
1 |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
UNITS = EV (1, UCUM, "no units") Value = Integer > = 1 |
|||
|
4 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) Linear Measurement shall be used. |
|||
|
5 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) Area Measurement shall be used. |
|||
|
6 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) Volume Measurement shall be used. |
This template provides the detail specific to a density.
Non-Extensible
Significant
Table TID 4011. Mammography CAD Density
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
||||
|
2 |
CODE |
EV (M-020F9, SRT, "Shape") |
1 |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
|||
|
3 |
CODE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
||||
|
4 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) Linear Measurement shall be used. |
|||
|
5 |
INCLUDE |
1-n |
MC |
At least one of rows 1, 2, 3, 4, 5 shall be present |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) Area Measurement shall be used. |
|||
|
6 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) Volume Measurement shall be used. |
This template provides the detail specific to a finding other than a lesion (see CID 6040 “Non-lesion Object Type”).
Non-Extensible
Significant
Table TID 4012. Mammography CAD Non-Lesion
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
|||||
|
2 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) Linear Measurement shall be used. |
||||
|
3 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) Area Measurement shall be used. |
||||
|
4 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) Volume Measurement shall be used. |
This template provides the detail specific to a selected region. A selected region is any mammography CAD derived arbitrary region of the image, whether within the breast outline or not. This can be use to delineate regions such as the intramammary fold.
Non-Extensible
Significant
Table TID 4013. Mammography CAD Selected Region
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
TEXT |
1 |
M |
|||||
|
2 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1400) Linear Measurement shall be used. |
||||
|
3 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1401) Area Measurement shall be used. |
||||
|
4 |
INCLUDE |
1-n |
U |
If the measured path is encoded, the SCOORD and its by-reference relationship to the IMAGE in TID (1402) Volume Measurement shall be used. |
This template provides the detail specific to image quality. It allows the encoding of descriptors of image quality (e.g., CID 6041 “Mammography Image Quality Finding”) for a given image or region of an image. For instance, images with partial motion blur can be identified with the region noted.
Table TID 4014. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$QualityFinding |
Coded term or Context Group for Quality Finding |
|
$QualityStandard |
Coded term or Context Group for Quality Control Standard |
Non-Extensible
Significant
Table TID 4014. CAD Image Quality
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
$QualityFinding |
||||
|
2 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
|
3 |
> |
HAS PROPERTIES |
CODE |
1 |
UC |
Shall be present if row 2 is present. |
$QualityStandard |
|
|
4 |
> |
HAS PROPERTIES |
NUM |
1 |
U |
UNITS = EV (%, UCUM, "Percent") Value = 0 - 100 |
Content Item Descriptions
|
Image Quality Rating |
A numeric value in the range 0 to 100, inclusive, where 0 is worst quality and 100 is best quality. |
This template gathers two lists of detection algorithms attempted, grouped by success/failure status. Algorithms not attempted are not mentioned in this sub-tree of the Document Root (e.g., TID 4000 “Mammography CAD Document Root”). This information forms the basis for understanding why a CAD report may produce no (or fewer than anticipated) detection results.
The sub-tree formed by this template is illustrated in Figure E.1-2 “Summary of Detections and Analyses Levels of Mammography CAD SR Content Tree” in PS3.17 .
Table TID 4015. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$DetectionCode |
Coded term or Context Group for Detection Performed |
Non-Extensible
Significant
Table TID 4015. CAD Detections Performed
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
MC |
Shall be present only if value of parent is (111222, DCM, "Succeeded") or (111223, DCM, "Partially Succeeded") |
||||
|
2 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$DetectionCode = $DetectionCode |
||
|
3 |
CONTAINER |
1 |
MC |
Shall be present only if value of parent is (111224, DCM, "Failed") or (111223, DCM, "Partially Succeeded") |
||||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$DetectionCode = $DetectionCode |
This template gathers two lists of analysis algorithms attempted, grouped by success/failure status. Algorithms not attempted are not mentioned in this sub-tree of the Document Root (e.g., TID 4000 “Mammography CAD Document Root”). This information forms the basis for understanding why a CAD report may produce no (or fewer than anticipated) analysis results.
The sub-tree formed by this template is illustrated in Figure E.1-2 “Summary of Detections and Analyses Levels of Mammography CAD SR Content Tree” in PS3.17 .
Table TID 4016. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$AnalysisCode |
Coded term or Context Group for Analysis Performed |
Non-Extensible
Significant
Table TID 4016. CAD Analyses Performed
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
MC |
Shall be present only if value of parent is (111222, DCM, "Succeeded") or (111223, DCM, "Partially Succeeded") |
||||
|
2 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$AnalysisCode = $AnalysisCode |
||
|
3 |
CONTAINER |
1 |
MC |
Shall be present only if value of parent is (111224, DCM, "Failed") or (111223, DCM, "Partially Succeeded") |
||||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$AnalysisCode = $AnalysisCode |
This template fully identifies a detection algorithm and the images and/or image regions on which it operated (see TID 4015 “CAD Detections Performed”).
Table TID 4017. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$DetectionCode |
Coded term or Context Group for Detection Performed |
Non-Extensible
Significant
Table TID 4017. CAD Detection Performed
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
$DetectionCode |
||||
|
2 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
M |
|||
|
3 |
> |
HAS PROPERTIES |
IMAGE |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
4 |
> |
R-HAS PROPERTIES |
IMAGE |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
Shall reference IMAGE content item(s) in the (111028, DCM, "Image Library") |
|
|
5 |
> |
HAS PROPERTIES |
UIDREF |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
6 |
> |
HAS PROPERTIES |
SCOORD |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
7 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR row 8 |
||
|
8 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR row 7 |
Shall reference an IMAGE content item in the (111028, DCM, "Image Library") |
|
|
9 |
> |
INCLUDE |
1 |
U |
Content Item Descriptions
|
CAD Algorithm Identification |
If more than one detection algorithm has the same "Detection Performed" code value (e.g., CID 6014 “Mammography Single Image Finding”) then the "CAD Algorithm Identification" shall unambiguously distinguish between algorithms. |
|
Rows 5, 6 |
Mammography CAD SR: When this template is invoked for the Mammography CAD SR, the Image Library is mandatory, thus only row 4 and/or row 6 shall be present. Chest CAD SR: When this template is invoked for the Chest CAD SR, the Image Library is optional, thus any combination of rows 3, 4, 5 and 6 may be present. Colon CAD SR: When this template is invoked for the Colon CAD SR, the Image Library does not exist, thus rows 3, 5, and/or 6 may be present and row 4 shall not be present. |
|
Rows 5, 6 |
Mammography CAD SR: When this template is invoked for the Mammography CAD SR, the Image Library is mandatory, thus only row 8 shall be present. Chest CAD SR: When this template is invoked for the Chest CAD SR, the Image Library is optional, thus row 7 or 8 may be present. Colon CAD SR: When this template is invoked for the Colon CAD SR, the Image Library does not exist, thus only row 7 may be present. |
This template fully identifies an analysis algorithm and the images and/or image regions on which it operated (see TID 4016 “CAD Analyses Performed”).
Table TID 4018. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$AnalysisCode |
Coded term or Context Group for Analysis Performed |
Non-Extensible
Significant
Table TID 4018. CAD Analysis Performed
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
$AnalysisCode |
||||
|
2 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
M |
|||
|
3 |
> |
HAS PROPERTIES |
IMAGE |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
4 |
> |
R-HAS PROPERTIES |
IMAGE |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
Shall reference IMAGE content item(s) in the (111028, DCM, "Image Library") |
|
|
5 |
> |
HAS PROPERTIES |
UIDREF |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
6 |
> |
HAS PROPERTIES |
SCOORD |
1-n |
MC |
At least one of row 3, 4, 5 or 6 shall be present |
||
|
7 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 8 |
||
|
8 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 7 |
Shall reference an IMAGE content item in the (111028, DCM, "Image Library") |
|
|
9 |
> |
INCLUDE |
1 |
U |
Content Item Descriptions
|
CAD Algorithm Identification |
If more than one analysis algorithm has the same "Analysis Performed" code value (e.g., CID 6043 “Types of Mammography CAD Analysis”) then the "CAD Algorithm Identification" shall unambiguously distinguish between algorithms. |
|
Rows 5, 6 |
Mammography CAD SR: When this template is invoked for the Mammography CAD SR, the Image Library is mandatory, and a total of at least two instances of row 4 or row 6 shall be present. Chest CAD SR: When this template is invoked for the Chest CAD SR, the Image Library is optional, thus any combination of rows 3, 4, 5 and 6 may be present. Colon CAD SR: When this template is invoked for the Colon CAD SR, the Image Library does not exist, thus rows 3, 5 and/or 6 may be present and row 4 shall not be present. |
|
Rows 5, 6 |
Mammography CAD SR: When this template is invoked for the Mammography CAD SR, the Image Library is mandatory, thus only row 8 shall be present. Chest CAD SR: When this template is invoked for the Chest CAD SR, the Image Library is optional, thus row 7 or 8 may be present. Colon CAD SR: When this template is invoked for the Colon CAD SR, the Image Library does not exist, thus only row 7 may be present. |
This template details the algorithm unambiguously. Re-state the software identification from the General Equipment Module of the SR IOD if all algorithms are unambiguously defined by that module.
Non-Extensible
Significant
Table TID 4019. CAD Algorithm Identification
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
TEXT |
1 |
M |
|||||
|
2 |
TEXT |
1 |
M |
|||||
|
3 |
TEXT |
1-n |
U |
Each instance of the Image Library Entry template contains the Image SOP Class and Instance UIDs, and selected attributes for an image that facilitate spatial analysis without having to retrieve the entire set of referenced images. If values for the attributes are not present in the Image SOP Instance, then as many of the attributes as possible should be derived.
Table TID 4020. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$ImageLaterality |
Coded term or Context Group for Image Laterality |
|
$ImageView |
Coded term or Context Group for Image View |
|
$ImageViewMod |
Coded term or Context Group for Image View Modifier |
Extensible
Significant
Table TID 4020. CAD Image Library Entry
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
IMAGE |
1 |
M |
|||||
|
2 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
MC |
Shall be present if (0020,0062) is in the Image Instance |
$ImageLaterality |
|
|
3 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
MC |
Shall be present if (0054,0220) is in the Image Instance |
$ImageView |
|
|
4 |
>> |
HAS CONCEPT MOD |
CODE |
1-n |
MC |
Shall be present if (0054,0222) is in the Image Instance |
$ImageViewMod |
|
|
5 |
> |
HAS ACQ CONTEXT |
TEXT |
1 |
MC |
Shall be present if (0020,0020) is in the Image Instance |
||
|
6 |
> |
HAS ACQ CONTEXT |
TEXT |
1 |
MC |
Shall be present if (0020,0020) is in the Image Instance |
||
|
7 |
> |
HAS ACQ CONTEXT |
DATE |
1 |
MC |
Shall be present if (0008,0020) is in the Image Instance |
||
|
8 |
> |
HAS ACQ CONTEXT |
TIME |
1 |
MC |
Shall be present if (0008,0030) is in the Image Instance |
||
|
9 |
> |
HAS ACQ CONTEXT |
DATE |
1 |
MC |
Shall be present if (0008,0023) is in the Image Instance |
||
|
10 |
> |
HAS ACQ CONTEXT |
TIME |
1 |
MC |
Shall be present if (0008,0033) is in the Image Instance |
||
|
11 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if (0018,1164) or (0028,0030) is in the Image Instance |
UNITS = EV (um, UCUM, "micrometer") UNITS = EV (mm, UCUM, "millimeter") |
|
|
12 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if (0018,1164) or (0028,0030) is in the Image Instance |
UNITS = EV (um, UCUM, "micrometer") UNITS = EV (mm, UCUM, "millimeter") |
|
|
13 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if (0018,1510) is in the Image Instance |
||
|
14 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if (0018,1511) is in the Image Instance |
||
|
15 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be computed from the Image Position (Patient) (0020,0032) projected onto the normal to the Image Orientation (Patient) (0020,0037) if present; may or may not be the same as the Spacing Between Slices (0018,0088) if present. |
UNITS = EV (mm, UCUM, "millimeter") |
|
|
16 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if Slice Thickness (0018,0050) is in the Image Instance. |
UNITS = EV (mm, UCUM, "millimeter") |
|
|
17 |
> |
HAS ACQ CONTEXT |
UIDREF |
1 |
UC |
May be present if Frame of Reference UID (0020,0052) is in the Image Instance. |
||
|
18 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if Image Position (Patient) (0020,0032) is in the Image Instance, and is the first value of Image Position (Patient) (0020,0032) for the referenced image or frame. |
UNITS = EV (mm, UCUM, "millimeter") |
|
|
19 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 18 is present, and is the second value of Image Position (Patient) (0020,0032) in the Image Instance for the referenced image or frame. |
UNITS = EV (mm, UCUM, "millimeter") |
|
|
20 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 18 is present, and is the second value of Image Position (Patient) (0020,0032) in the Image Instance for the referenced image or frame. |
UNITS = EV (mm, UCUM, "millimeter") |
|
|
21 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if Image Position (Patient) (0020,0037) is in the Image Instance, and is the first value of Image Orientation (Patient) (0020,0037) for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
22 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 21 is present, and is the second value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
23 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 21 is present, and is the third value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
24 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 21 is present, and is the fourth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
25 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 21 is present, and is the fifth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
26 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 21 is present, and is the sixth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame. |
UNITS = EV ({-1:1}, UCUM, "{-1:1}") |
|
|
27 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
UC |
May be present if Rows (0028,0010) is in the Image Instance. |
UNITS = EV ({pixels}, UCUM, "pixels") |
|
|
28 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
MC |
Shall be present if Row 27 is present, and is the value of Columns (0028,0011) in the Image Instance. |
UNITS = EV ({pixels}, UCUM, "pixels") |
Content Item Descriptions
|
Patient Orientation Row |
First (row) and second (column) components of Patient Orientation (0020,0020) in the Image IOD. See Section C.7.6.1.1.1 in PS3.3 . |
|
Patient Orientation Column |
|
|
Horizontal Imager Pixel Spacing |
The row (first) component of Imager Pixel Spacing (0018,1164) in the Image IOD. See Section C.8.11.4 in PS3.3 . |
|
Vertical Imager Pixel Spacing |
The column (second) component of Imager Pixel Spacing (0018,1164) in the Image IOD. See Section C.8.11.4 in PS3.3 . |
All geometry template invocations require specification of the location of the center of the object. Outline is optional.
Non-Extensible
Significant
Table TID 4021. Mammography CAD Geometry
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
SCOORD |
1 |
M |
GRAPHIC TYPE = {POINT} |
||||
|
2 |
> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Shall reference an IMAGE content item in the (111028, DCM, "Image Library") |
||
|
3 |
SCOORD |
1 |
U |
|||||
|
4 |
> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Shall reference the same content item as row 2 |
||
|
5 |
SCOORD |
1-n |
U |
|||||
|
6 |
> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Shall reference the same content item as row 2 |
This template is invoked when a content item, which may be the "root" of a sub-tree, is paraphrased from a prior SR document.
Non-Extensible
Significant
Table TID 4022. CAD Observation Context
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
COMPOSITE |
1 |
MC |
Shall be present if the original source is a DICOM object. |
||||
|
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
|
3 |
INCLUDE |
1 |
M |
This template describes CAD operating points. The description is deliberately left flexible and optional to allow implementation at differing levels of complexity.
Non-Extensible
Significant
Table TID 4023. CAD Operating Points
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
HAS PROPERTIES |
NUM |
1 |
M |
UNITS = DT ([arb'U], UCUM, "arbitrary unit") Value is restricted to being an integer |
|||
|
2 |
HAS PROPERTIES |
NUM |
1 |
U |
UNITS = DT ({0:n}, UCUM, "range: 0:n"), where n is the value specified in row 1 Value is restricted to being an integer |
|||
|
3 |
HAS PROPERTIES |
CONTAINER |
1 |
U |
||||
|
4 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
5 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
6 |
> |
CONTAINS |
NUM |
1-n |
M |
Number of instances of this row shall equal value of row 1, plus 1. |
UNITS = DT ({0:n}, UCUM, "range: 0:n"), where n is the value of Row 1. Value is restricted to being an integer that is unique within the invocation of this template. |
|
|
7 |
>> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
|
8 |
>> |
HAS PROPERTIES |
NUM |
The value of Row 4 |
1 |
U |
||
|
9 |
>> |
HAS PROPERTIES |
NUM |
The value of Row 5 |
1 |
U |
Content Item Descriptions
|
Maximum CAD Operating Point |
The maximum possible value of CAD Operating Point for this type of Detection Performed. No CAD Operating Point value recorded in the CAD Processing and Findings Summary sub-tree of the report for this type of Detection Performed shall exceed this value. The report may or may not contain Rendering Intent = "Presentation Optional" detections that are assigned the maximum value. |
|
Recommended CAD Operating Point |
A number indicating which of the CAD operating points is recommended by the creator of a CAD SR instance as the first operating point to be used when rendering the CAD SR instance contents. Subsequent changes to the displayed operating point are implementation dependent. |