Table A.3-1 specifies the Modules of the CT Image IOD.
If Multi-energy CT Acquisition (0018,9361) is YES the following constraints will apply:
The Contrast/Bolus Module shall be present if contrast was administered even if images are processed to remove contrast information from the pixels, e.g. Virtual Non-Contrast images.
The Real World Value Mapping Sequence (0040,9096) shall be present in the General Image Module.
For Measurement Units Code Sequence (0040,08EA) in the Real World Value Mapping Sequence (0040,9096) DCID 301 “Multi-energy Material Units” shall be used.