| DICOM PS3.3 2020a - Information Object Definitions |
|---|
The RT Radiation Common Module contains the Attributes shared by all RT Radiation IODs used for radiation treatment delivery.
Table C.36.13-1. RT Radiation Common Module Attributes
|
Include Table 10.9.1-1 “Enhanced Content Identification Macro Attributes”. |
|||
|
The level of detail of content within this SOP Instance. Enumerated Values:
|
|||
|
Whether or not device parameters about actual delivery of treatment to a patient have been recorded. |
|||
|
Only a single Item shall be included in this Sequence. See Section C.36.13.1.1. |
|||
|
Include Table C.36.2.2.4-1 “RT Treatment Position Macro Attributes”. |
See Section C.36.13.1.2. |
||
|
A set of tolerance values to be applied to parameters used for delivery of the RT Radiation. |
|||
|
>Include Table C.36.2.2.17-1 “RT Tolerance Set Macro Attributes”. |
|||
|
The maximum delivery time in seconds. See Section C.36.13.1.3. |
|||
|
A mode of operation on the treatment machine. Required if a special delivery mode is used for treatment. Only a single Item shall be included in this Sequence. See Section C.36.13.1.4. |
|||
The RT Treatment Technique Code Sequence (3010,0080) describes the treatment technique, i.e., how the radiation beam is shaped and targeted.
The RT Treatment Position Macro describes how the patient is to be positioned with respect to the delivery device for treatment.
The behavior of the machine regarding the patient position between specified Control Points is outside the scope of this Standard. Communicating devices shall agree on this behavior.
The Treatment Time Limit (300A,062E) is the maximum time span allowed to deliver a single fraction of this RT Radiation SOP instance. The behavior of the treatment delivery device on exceeding the Treatment Time Limit is up to the implementation. Treatment is expected to terminate upon reaching the Treatment Time Limit (300A,062E) independent of the Meterset and so this Attribute should have a value that accommodates normal variations in delivery.
The Treatment Machine Special Mode Code Sequence (300A,0635) contains a code, for example, (130341, DCM, "Total Body Irradiation") or (130342, DCM, "Total Skin Irradiation"), which selects a set of vendor- and machine-specific parameters that alter the treatment parameters and/or safety constraints.
| DICOM PS3.3 2020a - Information Object Definitions |
|---|