This Section describes Optical Surface Scanner Modules. These Modules contain Attributes that are specific to Optical Surface Scanner scans.
Table C.8.29-1 contains IOD Attributes that describe an optical surface scan series performed on the patient.
Table C.8.29-1. Optical Surface Scanner Series Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
Modality |
(0008,0060) |
1 |
Type of equipment that originally acquired the data used to create the images in this Series. Enumerated Values:
See Section C.7.3.1.1.1 for further explanation. |
|
Referenced Performed Procedure Step Sequence |
(0008,1111) |
1C |
Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related. Only a single Item shall be included in this sequence. Required if a Performed Procedure Step SOP Class was involved in the creation of this Series. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Referenced Surface Data Sequence |
(0080,0013) |
2 |
Lists the predecessor surfaces and/or point clouds instances. Zero or more Items shall be included in this sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||