Table C.8.12.7-1 specifies the Attributes that describe the Multi-Resolution Navigation Module, used in Whole Slide Microscopy LOCALIZER images.
Table C.8.12.7-1. Multi-Resolution Navigation Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
Referenced Image Navigation Sequence |
(0048,0200) |
1 |
Identification of correspondence between areas of Pixel Data (07FE,0010) and frames of higher resolution images. One or more Items shall be included in this sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
Referenced Image Instance |
||
|
>Referenced Frame Number |
(0008,1160) |
1 |
Referenced Frame Number within Referenced Image. Only a single value shall be present. |
|
>Top Left Hand Corner of Localizer Area |
(0048,0201) |
1 |
Top Left Hand Corner of Localizer image pixel area corresponding to referenced frame,specified as column\row. Column is the horizontal offset and row is the vertical offset. |
|
>Bottom Right Hand Corner of Localizer Area |
(0048,0202) |
1 |
Bottom Right Hand Corner of Localizer image pixel area corresponding to referenced frame,specified as column\row. Column is the horizontal offset and row is the vertical offset. |
|
>Pixel Spacing |
(0028,0030) |
1 |
Pixel Spacing in Referenced Image,specified as a numeric pair - adjacent row spacing \ adjacent column spacing, in mm. |
|
>Z Offset in Slide Coordinate System |
(0040,074A) |
1 |
The Z offset of the referenced frame from the Slide Coordinates System reference plane, in microns |
|
>Samples per Pixel |
(0028,0002) |
1 |
The number of samples (color planes) containing information in the referenced image. |
|
>Optical Path Identifier |
(0048,0106) |
1 |
Identifies the optical path used in acquiring the referenced Frame by reference to an Item in the Optical Path Sequence (0048,0105) of the current SOP Instance. See Section C.8.12.5. |