Table A.48-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Enhanced XRF Image IOD.
Table A.48-2. Enhanced XRF Image Functional Group Macros
|
Functional Group Macro |
Section |
Usage |
|---|---|---|
|
Frame Content |
M - May not be used as a Shared Functional Group. |
|
|
Referenced Image |
U |
|
|
Derivation Image |
C - Required if the image or frame has been derived from another SOP Instance. |
|
|
Cardiac Synchronization |
U |
|
|
Frame Anatomy |
M |
|
|
Frame VOI LUT |
M |
|
|
Contrast/Bolus Usage |
C - Required if the Enhanced Contrast/Bolus Module is present |
|
|
Pixel Intensity Relationship LUT |
C - Required if Pixel Intensity Relationship (0028,1040) equals LOG. May be present otherwise. |
|
|
Frame Pixel Shift |
U |
|
|
Patient Orientation in Frame |
U |
|
|
Frame Display Shutter |
U |
|
|
Respiratory Synchronization |
U |
|
|
Irradiation Event Identification |
M |
|
|
X-Ray Frame Characteristics |
U |
|
|
X-Ray Field of View |
U |
|
|
X-Ray Exposure Control Sensing Regions |
U |
|
|
X-Ray Frame Pixel Data Properties |
M |
|
|
X-Ray Frame Detector Parameters |
C - Required if X-Ray Receptor Type (0018,9420) is present and equals DIGITAL_DETECTOR. |
|
|
X-Ray Calibration Device Usage |
U |
|
|
X-Ray Object Thickness |
U |
|
|
X-Ray Frame Acquisition |
U |
|
|
X-Ray Positioner |
U |
|
|
X-Ray Table Position |
U |
|
|
X-Ray Collimator |
U |
|
|
X-Ray Geometry |
U |
The Defined CID for the Anatomic Region Sequence (0008,2218) shall be CID 4042 “XA/XRF Anatomy Imaged”.