Table 10-28 specifies the attributes that describe requirements on performing the movement of a device from an actual to a desired position.
Table 10-28. Device Motion Control Macro Attributes Description
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
Device Motion Control Sequence |
(300A,0450) |
3 |
Device Motion Control Definitions for each degree of freedom. One or more Items are permitted in this sequence. |
|
>Device Motion Parameter Code Sequence |
(300A,0453) |
1 |
The parameter for which the Device Motion Control shall be applied. |
|
>Device Motion Execution Mode |
(300A,0451) |
1C |
Permitted Motion Execution Mode for movement. Required if Device Motion Observation Mode (300A,0452) is absent. May be present otherwise. See Section 10.24.1.1. |
|
>Device Motion Observation Mode |
(300A,0452) |
1C |
Required Motion Observation Mode for movement. Required if Device Motion Observation Mode (300A,0452) is absent. May be present otherwise. See Section 10.24.1.2. |
Device Motion Execution Mode (300A,0451) identifies how the operator invokes and controls the motion.
Enumerated Values:
The device must be moved while the operator is activating one or more switches on the machine during the whole period of movement to prevent any uncontrolled movement.
The device can be moved automatically to the desired position, triggered by a one-time command of the operator, without the need to constantly activate any switch.
The device movement can be initiated and performed automatically by the device to the desired position.
Device Motion Observation Mode (300A,0452) describes the presence of the operator in the room, observing the motion.
Enumerated Values:
The movement of the device is only allowed when the operator is present in the treatment room.
The device can be moved by a command of the operator not being present in the treatment room, that is, from a device outside the treatment room.