Table C.8-45. RT General Plan Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
RT Plan Label |
(300A,0002) |
1 |
User-defined label for treatment plan. |
|
RT Plan Name |
(300A,0003) |
3 |
User-defined name for treatment plan. |
|
RT Plan Description |
(300A,0004) |
3 |
User-defined description of treatment plan. |
|
Instance Number |
(0020,0013) |
3 |
A number that identifies this object instance. |
|
RT Plan Date |
(300A,0006) |
2 |
Date treatment plan was last modified. |
|
RT Plan Time |
(300A,0007) |
2 |
Time treatment plan was last modified. |
|
Treatment Protocols |
(300A,0009) |
3 |
Planned treatment protocols. |
|
Plan Intent |
(300A,000A) |
3 |
Intent of this plan. Defined Terms:
|
|
Treatment Sites |
(300A,000B) |
3 |
Planned treatment sites. |
|
RT Plan Geometry |
(300A,000C) |
1 |
Describes whether RT Plan is based on patient or treatment device geometry. See Section C.8.8.9.1. Defined Terms:
|
|
Referenced Structure Set Sequence |
(300C,0060) |
1C |
The RT Structure Set on which the RT Plan is based. Only a single item shall be included in this sequence. Required if RT Plan Geometry (300A,000C) is PATIENT. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Referenced Dose Sequence |
(300C,0080) |
3 |
Related instances of RT Dose (for grids and named/unnamed point doses). One or more items are permitted in this sequence. See Note 1. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Referenced RT Plan Sequence |
(300C,0002) |
3 |
Related instances of RT Plan. One or more items are permitted in this sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
>RT Plan Relationship |
(300A,0055) |
1 |
Relationship of referenced plan with respect to current plan. Defined Terms:
|
An RT Dose IOD referenced within the Referenced Dose Sequence (300C,0080) can be used for storing grid-based (pixel) data, individual dose points (with optional dose point names), isodose curves, and DVHs.
An RT Plan Geometry (300A,000C) of PATIENT shall signify that an RT Structure Set has been defined upon which the plan geometry is based, and this RT Structure Set shall be specified in the Referenced Structure Set Sequence (300C,0060). An RT Plan Geometry (300A,000C) of TREATMENT_DEVICE shall indicate that no patient geometry is available, and that the RT Plan is being defined with respect to the IEC FIXED Coordinate System.