Table C.8.19.3-1 specifies the attributes of the XA/XRF Acquisition Module.
Table C.8.19.3-1. XA/XRF Acquisition Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
KVP |
(0018,0060) |
1 |
Average of the peak kilo voltage outputs of the X-Ray generator used for all frames. |
|
Radiation Setting |
(0018,1155) |
1 |
Identify the general level of X-Ray dose exposure. Enumerated Values:
|
|
X-Ray Tube Current in mA |
(0018,9330) |
1C |
Average of the nominal X-Ray tube currents in milliamperes for all frames. Required if Exposure in mAs (0018,9332) is not present. May be present otherwise. |
|
Exposure Time in ms |
(0018,9328) |
1C |
Duration of X-Ray exposure in milliseconds. See Section C.8.7.2.1.1. Required if Exposure in mAs (0018,9332) is not present. May be present otherwise. |
|
Exposure in mAs |
(0018,9332) |
1C |
The exposure expressed in milliampereseconds, for example calculated from Exposure Time and X-Ray Tube Current. Required if either Exposure Time in ms (0018,9328) or X-Ray Tube Current in mA (0018,9330) are not present. May be present otherwise. |
|
Average Pulse Width |
(0018,1154) |
1 |
Average width of X-Ray pulse in msec. |
|
Acquisition Duration |
(0018,9073) |
1 |
The time in seconds needed for the complete acquisition. See Section C.7.6.16.2.2.1 for further explanation |
|
Radiation Mode |
(0018,115A) |
1 |
Specifies X-Ray radiation mode. Defined Terms:
|
|
Focal Spot(s) |
(0018,1190) |
3 |
Nominal focal spot size in mm used to acquire this image. |
|
Anode Target Material |
(0018,1191) |
3 |
The primary material in the anode of the X-Ray source. Defined Terms:
|
|
Rectification Type |
(0018,1156) |
3 |
Type of rectification used in the X-Ray generator. Defined Terms:
|
|
X-Ray Receptor Type |
(0018,9420) |
1 |
Identifies with type of X-Ray receptor is used. Enumerated Values:
|
|
Distance Receptor Plane to Detector Housing |
(0018,9426) |
2 |
Distance in mm between the receptor plane and the detector housing. The direction of the distance is positive from receptor plane to X-Ray source. Note
|
|
Positioner Type |
(0018,1508) |
1 |
Defined Terms:
Note
|
|
C-arm Positioner Tabletop Relationship |
(0018,9474) |
1C |
Describes for C-arm positioner type systems if positioner and tabletop has the same geometrical reference system. Enumerated Values:
NoteThe value NO is intended for mobile systems where there is no table fixed to the system Required if Positioner Type (0018,1508) equals CARM. |
|
Acquired Image Area Dose Product |
(0018,9473) |
2 |
X-Ray dose, measured in dGy*cm*cm, to which the patient was exposed for the acquisition of this image only. Note
|