This template collects a composite feature for a lesion, non-lesion object, or correlation of related objects. The details of the composition are expressed in the Mammography CAD Composite Feature Body (see TID 4005 “Mammography CAD Composite Feature Body”). The data from which the details are inferred, are expressed in the Composite Features (see TID 4004 “Mammography CAD Composite Feature”) and/or Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”), of which there may be several.
A Composite Feature shall be INFERRED FROM any combination of two or more Composite Features or Single Image Findings or mixture thereof.
Non-Extensible
Significant
Table TID 4004. Mammography CAD Composite Feature
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
|
4 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
M |
|||
|
5 |
> |
INFERRED FROM |
INCLUDE |
1-n |
MC |
At least two items shall be present: two of row 5, two of row 6, or one of each. |
||
|
6 |
> |
INFERRED FROM |
INCLUDE |
1-n |
MC |
At least two items shall be present: two of row 5, two of row 6, or one of each. |
||
|
7 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
Shall be present only if this feature is incorporated from a different report than its parent. |
Content Item Descriptions
|
Rendering Intent |
This content item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this template and its target content items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 . |