The templates that comprise the CT Radiation Dose SR are interconnected as in Figure A-15.
This template defines a container (the root) with subsidiary content items, each of which corresponds to a single CT X-Ray irradiation event entry. There is a defined recording observer (the system or person responsible for recording the log, generally the system). Accumulated values shall be kept for a whole Study or at least a part of a Study, if the Study is divided in the workflow of the examination, or a performed procedure step. Multiple CT Radiation Dose objects may be created for one Study.
Extensible
Significant
Yes
Table TID 10011. CT Radiation Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
Root node |
||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
EV (P5-08000, SRT, "Computed Tomography X-Ray") |
||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E8, SRT, "Has Intent") |
1 |
M |
||
|
4 |
> |
INCLUDE |
1-n |
M |
||||
|
5 |
> |
HAS OBS CONTEXT |
DATETIME |
1 |
M |
|||
|
6 |
> |
HAS OBS CONTEXT |
DATETIME |
1 |
M |
|||
|
7 |
> |
HAS OBS CONTEXT |
CODE |
1 |
M |
|||
|
8 |
>> |
HAS PROPERTIES |
UIDREF |
1 |
M |
|||
|
9 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
10 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
|||
|
11 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
12 |
> |
CONTAINS |
CODE |
1-n |
M |
|||
|
13 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
Content Item Descriptions
|
Row 4 |
The observer context may include both a Person Observer identification, as well as the identity of the equipment providing the values for the irradiation event (Device Observer identification), if not inherited. |
|
Row 5 |
Start, Date Time of the first CT Irradiation Event of the accumulation |
|
Row 6 |
End, Date Time of the last CT Irradiation Event of the accumulation |
|
Row 12 |
The primary source of information from which this dose object was constructed. |
|
Row 13 |
The physician responsible for determining that the irradiating procedure was appropriate for the indications. The value may come from Requesting Physician (0032,1032), Requesting Physician Identification Sequence (0032,1031) or somewhere else based on hospital policies. |