This section describes the specific modules for the Enhanced MR Image IOD.
This section describes the Enhanced MR Image Module.
Table C.8-79 specifies the attributes of the Enhanced MR Image module.
Table C.8-79. Enhanced MR Image Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
Include Table C.8-83 “MR Image and Spectroscopy Instance Macro Attributes” |
|||
|
Image Type |
(0008,0008) |
1 |
Image characteristics. See Section C.8.16.1 and Section C.8.13.1.1.1. |
|
Include Table C.8-131 “Common CT/MR Image Description Macro Attributes” |
|||
|
Include Table C.8-84 “MR Image Description Macro Attributes” |
|||
|
Samples per Pixel |
(0028,0002) |
1 |
Number of samples (planes) in this image. For Enumerated Values See Section C.8.13.1.1.2. |
|
Photometric Interpretation |
(0028,0004) |
1 |
Specifies the intended interpretation of the pixel data. Enumerated Values are specified in the IOD that invokes this Module. See Section C.7.6.3.1.2 for definition of this term. |
|
Bits Allocated |
(0028,0100) |
1 |
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. For Enumerated Values See Section C.8.13.1.1.2. |
|
Bits Stored |
(0028,0101) |
1 |
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. For Enumerated Values See Section C.8.13.1.1.2. |
|
High Bit |
(0028,0102) |
1 |
Most significant bit for pixel sample data. Each sample shall have the same high bit. Shall be one less than the value in Bits Stored (0028,0101). |
|
Pixel Representation |
(0028,0103) |
1 |
Data representation of the pixel samples. Each sample shall have the same pixel representation. For Enumerated Values See Section C.8.13.1.1.2 |
|
Planar Configuration |
(0028,0006) |
1C |
Indicates whether the pixel data are sent color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a value greater than 1. See Section C.7.6.3.1.3 and Section C.8.13.1.1.2 for further explanation. |
|
Spacing Between Slices |
(0018,0088) |
3 |
Value of the prescribed spacing to be applied between the slices in a volume that is to be acquired. The spacing in mm is defined as the center-to-center distance of adjacent slices. |
|
Burned In Annotation |
(0028,0301) |
1C |
Indicates whether or not the image contains sufficient burned in annotation to identify the patient and date the image was acquired. Enumerated Values:
This means that images that contain this Module shall not contain such burned in annotations. Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted), may be present otherwise. |
|
Recognizable Visual Features |
(0028,0302) |
3 |
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. Enumerated Values:
If this Attribute is absent, then the image may or may not contain recognizable visual features. |
|
Lossy Image Compression |
(0028,2110) |
1C |
Specifies whether an Image has undergone lossy compression (at a point in its lifetime). Enumerated Values:
Once this value has been set to 01 it shall not be reset. See Section C.7.6.1.1.5. Required if SOP Class UID is not "1.2.840.10008.5.1.4.1.1.4.4" (Legacy Converted). May be present otherwise. |
|
Lossy Image Compression Ratio |
(0028,2112) |
1C |
Describes the approximate lossy compression ratio(s) that have been applied to this image. See Section C.7.6.1.1.5 for further explanation. May be multi-valued if successive lossy compression steps have been applied; the value order shall correspond to the values of Lossy Image Compression Method (0028,2114). Note
Required if Lossy Image Compression (0028,2110) is "01". |
|
Lossy Image Compression Method |
(0028,2114) |
1C |
A label for the lossy compression method(s) that have been applied to this image. See Section C.7.6.1.1.5 for further explanation. May be multi-valued if successive lossy compression steps have been applied; the value order shall correspond to the values of Lossy Image Compression Ratio (0028,2112). NoteFor historical reasons, the lossy compression method may also be described in Derivation Description (0008,2111). Required if Lossy Image Compression (0028,2110) is "01". |
|
Presentation LUT Shape |
(2050,0020) |
1C |
Specifies an identity transformation for the Presentation LUT, such that the output of all grayscale transformations defined in the IOD containing this Module are defined to be P-Values. Enumerated Values:
Required if Photometric Interpretation (0028,0004) is MONOCHROME2. |
|
Icon Image Sequence |
(0088,0200) |
3 |
This icon image is representative of the Image. Only a single Item is permitted in this Sequence. |
|
See Section C.7.6.1.1.6. |
|||
|
Include Table 10-25 “Optional View and Slice Progression Direction Attributes” |
|||
Image Type (0008,0008) and Frame Type (0008,9007) are not included in this Macro but one or the other is always included in the Module or Macro that invokes this Macro, and they are therefore described here.
In addition to the requirements specified in Section C.8.16.1 Image Type and Frame Type, the following additional requirements and Defined Terms are specified.
Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.1. No additional requirements or Defined Terms.
Value 2 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.2. No additional requirements or Defined Terms.
Table C.8-80 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.3 for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-80. MR-Specific Image Type and Frame Type Value 3
|
Defined Term Name |
Defined Term Description |
|---|---|
|
ANGIO_TIME |
Angio time acquisition (peripheral vascular/carotid) |
|
ASL |
Arterial Spin Labeling |
|
CINE |
Cardiac CINE |
|
DIFFUSION |
Collected to show diffusion effects. |
|
DIXON |
Dixon Water Fat Imaging Techniques. |
|
FLOW_ENCODED |
Flow Encoded |
|
FLUID_ATTENUATED |
Fluid Attenuated T2 weighted |
|
FMRI |
Collected for functional imaging calculations. |
|
MAX_IP |
Maximum Intensity Projection |
|
MIN_IP |
Minimum Intensity Projection |
|
M_MODE |
Image line over time |
|
METABOLITE_MAP |
Metabolite Maps from spectroscopy data |
|
MULTIECHO |
Multiple echoes with different contrast weighting (e.g., proton density and T2 weighted) |
|
PROTON_DENSITY |
Proton density weighted |
|
REALTIME |
Real-time collection of single slices |
|
STIR |
Short Tau Inversion Recovery |
|
TAGGING |
Images with superposition of thin saturation bands |
|
TEMPERATURE |
Images record temperature |
|
T1 |
T1 weighted |
|
T2 |
T2 weighted |
|
T2_STAR |
T2* weighted |
|
TOF |
Time Of Flight weighted |
|
VELOCITY |
Velocity encoded |
Table C.8-81 specifies the Defined Terms for MR additional to those defined in Section C.8.16.1.4 for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-81. MR-Specific Image Type and Frame Type Value 4
|
Defined Term Name |
Defined Term Description |
|---|---|
|
ADC |
Apparent Diffusion Coefficient |
|
DIFFUSION |
Diffusion weighted |
|
DIFFUSION_ANISO |
Diffusion Anisotropy |
|
DIFFUSION_ATTNTD |
Diffusion Attenuated. Derived by removing the T2 contributions from a Diffusion Weighted image. |
|
DIFFUSION_ISO |
Isotropic images derived from Directional Diffusion images |
|
FAT |
Fat images derived using Dixon or other techniques |
|
FAT_FRACTION |
Fat Fraction images derived using Dixon or other techniques |
|
FIELD_MAP |
Field Map images derived using Dixon or other techniques |
|
IN_PHASE |
Water/Fat In Phase images derived using Dixon or other techniques |
|
METABOLITE_MAP |
Metabolite Maps from spectroscopy data |
|
NEI |
Created through Negative Enhancement Integral operation |
|
OUT_OF_PHASE |
Water/Fat Out of phase images derived using Dixon or other techniques |
|
PERFUSION_ASL |
Perfusion from an ASL series obtained by subtraction of control and label data |
|
R_COEFFICIENT |
R-Coefficient Map (fMRI) |
|
R2_MAP |
R2 Map |
|
R2_STAR_MAP |
R2* Map images derived using Dixon or other techniques |
|
RHO |
Proton Density map |
|
SCM |
Signal Change Map |
|
SNR_MAP |
Signal to Noise Map |
|
T1_MAP |
T1 Map |
|
T2_STAR_MAP |
T2* Map images derived using Dixon or other techniques |
|
T2_MAP |
T2 Map |
|
TCS |
Time Course of Signal |
|
TEMPERATURE |
Temperature encoded |
|
VELOCITY |
Velocity encoded |
|
WATER |
Water images derived using Dixon or other techniques |
|
WATER_FRACTION |
Water fraction images derived using Dixon or other techniques |
Table C.8-82 specifies the Enumerated Values and allowed combinations of Samples per Pixel (0028,0002), Planar Configuration (0028,0006), Pixel Representation (0028,0103), Bits Allocated (0028,0100) and Bits Stored (0028,0101) for each allowable Photometric Interpretation allowed by the IOD that invokes this Module.
Table C.8-82. Allowed Combinations of Attribute Values for Photometric Interpretation, Samples Per Pixel, Planar Configuration, Pixel Representation, Bits Allocated and Bits Stored
|
Photometric Interpretation |
Samples per Pixel |
Planar Configuration |
Pixel Representation |
Bits Allocated |
Bits Stored |
|---|---|---|---|---|---|
|
MONOCHROME2 |
1 |
- |
0 or 1 |
8 |
8 |
|
MONOCHROME2 |
1 |
- |
0 or 1 |
16 |
12, 16 |
|
RGB |
3 |
0 |
0 |
8 |
8 |
|
YBR_ICT |
3 |
0 |
0 |
8 |
8 |
|
YBR_RCT |
3 |
0 |
0 |
8 |
8 |
|
YBR_PARTIAL_420 |
3 |
0 |
0 |
8 |
8 |
|
YBR_FULL_422 |
3 |
0 |
0 |
8 |
8 |
See Section C.8.16.2.1.1. No additional requirements or Defined Terms.
See Section C.8.16.2.1.2. No additional requirements or Defined Terms.
See Section C.8.16.2.1.3. No additional requirements or Defined Terms.