Table A.36-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Enhanced MR Image IOD.
Table A.36-2. Enhanced MR Image Functional Group Macros
|
Functional Group Macro |
Section |
Usage |
|---|---|---|
|
Pixel Measures |
M |
|
|
Frame Content |
M - May not be used as a Shared Functional Group. |
|
|
Plane Position (Patient) |
M |
|
|
Plane Orientation (Patient) |
M |
|
|
Referenced Image |
C - Required if the image or frame has been planned on another image or frame. May be present otherwise |
|
|
Derivation Image |
C - Required if the image or frame has been derived from another SOP Instance. |
|
|
Cardiac Synchronization |
C - Required if Cardiac Synchronization Technique (0018,9037) equals other than NONE and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
Frame Anatomy |
M |
|
|
Pixel Value Transformation |
C - Required if Photometric Interpretation (0028,0004) is MONOCHROME2. |
|
|
Frame VOI LUT |
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2. |
|
|
Real World Value Mapping |
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2. |
|
|
Contrast/Bolus Usage |
C - Required if Contrast/Bolus Agent Sequence (0018,0012) is used. |
|
|
Respiratory Synchronization |
C - Required if Respiratory Motion Compensation Technique (0018,9170) equals other than NONE, REALTIME or BREATH_HOLD and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Image Frame Type |
M |
|
|
MR Timing and Related Parameters |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR FOV/Geometry |
C - Required if Geometry of k-Space Traversal (0018,9032) equals RECTILINEAR and if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Echo |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Modifier |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Imaging Modifier |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Receive Coil |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Transmit Coil |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Diffusion |
C - Required if Acquisition Contrast (0008,9209) in any MR Image Frame Type Functional Group in the SOP Instance equals DIFFUSION and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Averages |
C - Required if Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Spatial Saturation |
C - Required if Spatial Pre-saturation (0018,9027) equals SLAB for any frame in the SOP Instance and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Metabolite Map |
C - Required if Image Type (0008,0008) Value 3 equals METABOLITE_MAP. May be present otherwise. |
|
|
MR Velocity Encoding |
C - Required if Phase Contrast (0018,9014) equals YES and Image Type (0008,0008) Value 1 is ORIGINAL or MIXED. May be present otherwise. |
|
|
MR Arterial Spin Labeling |
C - Required if Image Type (0008,0008) Value 3 is ASL. May be present otherwise. |