Table A.35.4-1 specifies the Modules of the Key Object Selection Document IOD.
Table A.35.4-1. Key Object Selection Document IOD Modules
|
IE |
Module |
Reference |
Usage |
|---|---|---|---|
|
Patient |
Patient |
M |
|
|
Clinical Trial Subject |
U |
||
|
Study |
General Study |
M |
|
|
Patient Study |
U |
||
|
Clinical Trial Study |
U |
||
|
Series |
Key Object Document Series |
M |
|
|
Clinical Trial Series |
U |
||
|
Equipment |
General Equipment |
M |
|
|
Document |
Key Object Document |
M |
|
|
SR Document Content |
M |
||
|
SOP Common |
M |
The Specimen Identification Module was previously included in this IOD but has been retired. See PS3.3-2008.
Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):
Enumerated Values:
Relationships between Content Items in the content of this IOD shall be conveyed in the by-value mode. See Table C.17.3-8 for Relationship Type definitions.
Relationships by-reference are forbidden. Therefore, Referenced Content Item Identifier (0040,DB73) is not present in any of the Content Items within the SR Document Content Module.
Table A.35.4-2 specifies the relationship constraints of this IOD.
Table A.35.4-2. Relationship Content Constraints for Key Object Selection Document IOD
|
Source Value Type |
Relationship Type (Enumerated Values) |
Target Value Type |
|---|---|---|
|
CONTAINER |
CONTAINS |
TEXT, IMAGE, WAVEFORM, COMPOSITE |
|
CONTAINER |
HAS OBS CONTEXT |
TEXT, CODE, UIDREF, PNAME |
|
CONTAINER |
HAS CONCEPT MOD |
CODE |
The document shall be constructed from TID 2010 “Key Object Selection” invoked at the root node.