This Application Profile is based on the Media Storage Service Class (see PS3.4).
Table D.3-1. STD-GEN SOP Classes and Transfer Syntaxes
|
Information Object Definition |
SOP Class UID |
Transfer Syntax and UID |
FSC Requirement |
FSR Requirement |
FSU Requirement |
|---|---|---|---|---|---|
|
Basic Directory |
1.2.840.10008.1.3.10 |
Explicit VR Little Endian Uncompressed 1.2.840.10008.1.2.1 |
Mandatory |
Mandatory |
Mandatory |
|
Composite Image & Stand-alone Storage |
See PS3.4 |
Explicit VR Little Endian Uncompressed 1.2.840.10008.1.2.1 |
Defined in Conformance Statement |
Defined in Conformance Statement |
Optional |
The SOP Classes and corresponding Transfer Syntax supported by this Application Profile are specified in the Table D.3-1. The supported Storage SOP Class(es) shall be listed in the Conformance Statement using a table of the same form.
The STD-GEN-CD and STD-GEN-SEC-CD application profiles require the 120 mm CD-R physical medium with the ISO/IEC 9660 Media Format, as defined in PS3.12.
The STD-GEN-DVD-RAM and STD-GEN-SEC-DVD-RAM application profiles require the 120 mm DVD-RAM medium, as defined in PS3.12.
The STD-GEN-BD and STD-GEN-SEC-BD application profiles require any of the 120 mm BD media, as defined in PS3.12.
Conformant Application Entities shall include in the DICOMDIR File the Basic Directory IOD containing Directory Records at the Patient and the subsidiary Study and Series levels, appropriate to the SOP Classes in the File Set.
All DICOM files in the File Set incorporating SOP Instances defined for the specific Application Profile shall be referenced by Directory Records.
DICOMDIRs with no directory information are not allowed by this Application Profile.
All implementations shall include the DICOM Media Storage Directory in the DICOMDIR file. There shall only be one DICOMDIR file per File Set. The DICOMDIR file shall be in the root directory of the medium. The Patient ID at the patient level shall be unique for each patient directory record in one File Set.
File Set Creators and Updaters are required to generate the mandatory elements specified in PS3.3.
Table D.3-2 specifies the additional associated keys. At each directory record level other additional data elements can be added, but it is not required that File Set Readers be able to use them as keys. Refer to the Basic Directory IOD in PS3.3.
Table D.3-2. STD-GEN Additional DICOMDIR Keys
|
Key Attribute |
Tag |
Directory Record Type |
Type |
Notes |
|---|---|---|---|---|
|
Image Type |
(0008,0008) |
IMAGE |
1C |
Required if present in image object. |
|
Referenced Image Sequence |
(0008,1140) |
IMAGE |
1C |
Required if present in image object. |
|
>Referenced SOP Class UID |
(0008,1150) |
IMAGE |
1C |
Required if Referenced Image Sequence (0008,1140) is present |
|
>Referenced SOP Instance UID |
(0008,1155) |
IMAGE |
1C |
Required if Referenced Image Sequence (0008,1140) is present. |
|
>All other elements from Referenced Image Sequence (including Purpose of Reference Code Sequence and its content) |
IMAGE |
3 |
||
The requirements with respect to the mandatory DICOMDIR keys in PS3.3 imply that either these attributes are present in the Image IOD, or they are in some other way supplied by the File-set Creator. These attributes are (0010,0020) Patient ID, (0008,0020) Study Date, (0008,0030) Study Time, (0020,0010) Study ID, (0020,0011) Series Number, and (0020,0013) Instance Number.
The STD-GEN-SEC-CD, STD-GEN-SEC-DVD-RAM and STD-GEN-SEC-BD application profiles require that all DICOM Files in the File-set including the DICOMDIR be Secure DICOM Files encapsulated in accordance with the requirements of the Basic DICOM Media Security Profile as defined in PS3.15.
These Application Profiles do not place any consistency restrictions on the use of the Basic DICOM Media Security Profile with different DICOM Files of one File-set. For example, readers should not assume that all Files in the File-set can be decoded by the same set of recipients. Readers should also not assume that all secure Files use the same approach (hash key or digital signature) to ensure Integrity or carry the same originators' signatures.