The templates that comprise the Quantitative Arterial Analysis SR are interconnected as in Figure A-4:
The Quantitative Arterial Analysis Template consists of a CONTAINER with a structure for reporting the result of the quantitative arterial analysis process.
Extensible
Significant
Table TID 3213. Quantitative Arterial Analysis
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
|
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1-n |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
5 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
6 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
7 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
|
8 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
Content Item Descriptions
|
Row 5 |
Identifies the Arterial Analysis program |
|
Row 6 |
Identifies the Arterial Analysis program version |
|
Row 7 |
Identifies the Arterial Analysis program manufacturer |
The Analyzed Segment Template consists of a CONTAINER providing quantitative arterial analysis measurements derived from the angiographic images.
Extensible
Significant
Table TID 3214. Analyzed Segment
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
3 |
> |
CONTAINS |
IMAGE |
1 |
M |
|||
|
4 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
5 |
> |
HAS ACQ CONTEXT |
INCLUDE |
1 |
U |
|||
|
6 |
> |
HAS ACQ CONTEXT |
CODE |
EV (G-72BB, SRT, "Catheterization Procedure Phase") |
1 |
U |
||
|
7 |
> |
CONTAINS |
SCOORD |
1 |
M |
GRAPHIC TYPE = POLYLINE |
||
|
8 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Must reference Row 3 |
||
|
9 |
> |
CONTAINS |
SCOORD |
1 |
M |
GRAPHIC TYPE = POLYLINE |
||
|
10 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
M |
Must reference Row 3 |
||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
12 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm, UCUM, "mm") |
||
|
13 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (G-A437, SRT, "Maximum") $Unit = DT (mm, UCUM, "mm") |
||
|
14 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
|
15 |
>> |
CONTAINS |
NUM |
1 |
M |
Value = 1 UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
16 |
>> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Unit = DT (mm, UCUM, "mm") |
||
|
17 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
18 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
19 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
|
20 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
|
21 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 1 |
Observation DateTime (0040,A032) of container needs to be flagged with the time of the analysis |
|
Row 6 |
Numeric coordinates (x,y) identifying the contour points from proximal to distal of left contour. Left is relative to the direction of the blood flow. |
|
Row 9 |
Numeric coordinates (x,y) identifying the contour points from proximal to distal of right contour. Right is relative to the direction of the blood flow. |
|
Row 12, 13 |
Positions are relative to the midpoint between the first left and right contour points and measured along the midline between the left and right contour. |
|
Row 14 |
The X-axis represents the pixel points of the midline of the vessel from proximal to distal. The points on the midline are not necessarily equidistant. |
|
Row 16 |
For each point of the midline of the vessel a measurement value for the diameter is calculated. |
|
Row 17, 18 |
The positions in the graph are related to the points on the midline of the vessel. |
|
Row 21 |
Secondary Capture image with Arterial Analysis contour. |
Definition of Left and Right defined by the direction of the blood flow as in Figure A-4b:
The Angiographic Lesion Analysis Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.
Extensible
Significant
Table TID 3215. Angiographic Lesion Analysis
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
EV (F-00585, SRT, "Lesion Finding") |
1 |
M |
||||
|
2 |
> |
CONTAINS |
TEXT |
1 |
M |
|||
|
3 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
4 |
>>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
|
5 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm, UCUM, "mm") |
||
|
6 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm2, UCUM, "mm2") |
||
|
7 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
8 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
|
9 |
>> |
CONTAINS |
NUM |
1-n |
M |
UNITS = DT (mm, UCUM, "mm") |
||
|
10 |
>>> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Unit = DT (mm, UCUM, "mm") |
||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $TargetSite = (122382, DCM, "Site of Luminal Minimum") $Unit = DT (mm, UCUM, "mm") |
||
|
12 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (122404, DCM, "Reconstructed") $TargetSite = (122382, DCM, "Site of Luminal Minimum") $Unit = DT (mm2, UCUM, "mm2") |
||
|
13 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-41D2D, SRT, "Calculated") $TargetSite = EV (122481, DCM, "Contour Start") $Unit = DT (mm, UCUM, "mm") |
||
|
14 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-41D2D, SRT, "Calculated") $TargetSite = EV (122482, DCM, "Contour End") $Unit = DT (mm, UCUM, "mm") |
||
|
15 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
16 |
> |
CONTAINS |
CONTAINER |
EV (122517, DCM, "Densitometric Luminal Cross-sectional Area Graph") |
1 |
U |
||
|
17 |
>> |
CONTAINS |
NUM |
1 |
M |
Value = 1 UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
18 |
>> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Unit = (mm2, UCUM, "mm2") |
||
|
19 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (R-41D2D, SRT, "Calculated") $Method = EV (122474, DCM, "Densitometric method") $TargetSite = EV (122481, DCM, "Contour Start") $Unit = (mm2, UCUM, "mm2") |
||
|
20 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (R-41D2D, SRT, "Calculated") $Method = EV (122474, DCM, "Densitometric method") $TargetSite = EV (122482, DCM, "Contour End") $Unit = (mm2, UCUM, "mm2") |
||
|
21 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (R-101BC, SRT, "Stenotic Lesion Length") $Unit = DT (mm, UCUM, "mm") |
||
|
22 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (R-101BB, SRT, "Lumen Diameter Stenosis") $Unit = DT (%, UCUM, "%") |
||
|
23 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (R-101BA, SRT, "Lumen Area Stenosis") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Unit = DT (%, UCUM, "%") |
||
|
24 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (122372, DCM, "Lumen Volume") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Unit = DT (mm3, UCUM, "mm3") |
||
|
25 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122542, DCM, "Plaque Area") $Unit = DT (mm2, UCUM, "mm2") |
||
|
26 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122376, DCM, "Total Plaque Volume") $Unit = DT (mm3, UCUM, "mm3") |
||
|
27 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122544, DCM, "Diameter Symmetry") $Unit = DT ({ratio}, UCUM, "ratio") |
||
|
28 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122545, DCM, "Area Symmetry") $Unit = DT ({ratio}, UCUM, "ratio") |
||
|
29 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122546, DCM, "Inflow Angle") $Unit = DT (deg, UCUM, "deg") |
||
|
30 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122547, DCM, "Outflow Angle") $Unit = DT (deg, UCUM, "deg") |
||
|
31 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
32 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 8 |
Set of user defined reference position for method that requires local reference position. |
|
Row 9 |
Distance of local reference position from an arbitrary landmark. |
|
Row 10 |
Diameter at a local reference position. |
|
Row 11 |
The reference diameter for the arterial lesion calculated with the applicable reference method |
|
Row 12 |
The reference area for the arterial lesion calculated with the applicable reference method |
|
Row 13 |
The diameter measurement at the start of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14) |
|
Row 14 |
The diameter measurement at the end of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14) |
|
Row 15 |
The positions of the lesion, borders of the lesion, etc. |
|
Row 16 |
The graph with the calculated cross sectional area results based on the densitometric method |
|
Row 18 |
The cross sectional area measurements calculated based on the densitometric method |
|
Row 19 |
The cross sectional area measurement at the start of the reconstruction line in the area graph |
|
Row 20 |
The cross sectional area measurement at the end of the reconstruction line in the area graph |
|
Row 21 |
Measured along the midline of the left and right contour |
|
Row 22 |
The diameter stenosis is calculated as follows: (Reference Luminal Diameter - Minimum Luminal Diameter / Reference Luminal Diameter) * 100% |
|
Row 23 |
The circular and the densitometric area stenosis are calculated respectively as: (Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Circular Area / Reference Vessel Lumen Cross-Sectional Area) * 100%(Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Densitometric Area / Reference Vessel Lumen Cross-Sectional Area) * 100% |
|
Row 24 |
Estimated lumen volume between proximal border and distal border of lesion (TID 3218 “Position in Arterial Segment”, row 1 and 2) |
|
Row 32 |
Secondary Capture image with obstruction analysis contour |
The Stenotic Flow Reserve Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.
Extensible
Significant
Table TID 3216. Stenotic Flow Reserve
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122548, DCM, "Stenotic Flow Reserve") $Unit = DT ({ratio}, UCUM, "ratio") |
|||
|
2 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122549, DCM, "Poiseuille Resistance") $Unit = DT (mm[Hg].s/cm, UCUM, "mmHG.s/cm") |
|||
|
3 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122550, DCM, "Turbulence Resistance") $Unit = DT (mm[Hg].s2/cm2, UCUM, "mmHG.s2/cm2") |
|||
|
4 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122555, DCM, "Estimated Normal Flow") $Unit = DT (ml/s, UCUM, "ml/s") |
|||
|
5 |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122551, DCM, "Pressure Drop at SFR") $Unit = DT (mm[Hg], UCUM, "mmHg") |
|||
|
6 |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 6 |
Secondary Capture image with SFR analysis contour |
The Sub-segmental Data Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for a sub-segment in a total analyzed segment.
Extensible
Significant
Table TID 3217. Sub-Segmental Data
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
|
4 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
5 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
6 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
7 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 7 |
Secondary Capture image with obstruction analysis contour |
The Position in Arterial Segment Template consists of the position content items common for the Angiographic Lesion Analysis and Sub-Segmental Data.
Extensible
Significant
Table TID 3218. Position in Arterial Segment
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122528, DCM, "Position of Proximal Border") $Unit = DT (mm, UCUM, "mm") |
|||
|
2 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122529, DCM, "Position of Distal Border") $Unit = DT (mm, UCUM, "mm") |
|||
|
3 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122382, DCM, "Site of Luminal Minimum") $Unit = DT (mm, UCUM, "mm") |
|||
|
4 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122516, DCM, "Site of Luminal Maximum") $Unit = DT (mm, UCUM, "mm") |
|||
|
5 |
CONTAINS |
NUM |
1 |
UC |
IFF TID 3214 “Analyzed Segment” Row 14 is present |
UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
6 |
CONTAINS |
NUM |
1 |
UC |
IFF TID 3214 “Analyzed Segment” Row 14 is present |
UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
7 |
CONTAINS |
NUM |
1 |
UC |
IFF TID 3214 “Analyzed Segment” Row 14 is present |
UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
8 |
CONTAINS |
NUM |
1 |
UC |
IFF TID 3214 “Analyzed Segment” Row 14 is present |
UNITS = DT ({pixels}, UCUM, "pixels") |
Content Item Descriptions
|
Row 1, 2, 3, 4 |
Positions are relative to the midpoint of the first left and right contour points and measured along the midline of the left and right contour |
|
Row 5, 6, 7, 8 |
The positions are relative to the measurement locations of the Diameter Graph of TID 3214 “Analyzed Segment” row 14. |
The Segment Values Template consists of content items providing quantitative arterial analysis measurements for a total analyzed segment or sub segment.
Extensible
Significant
Table TID 3219. Segment Values
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (122510, DCM, "Length Luminal Segment") $Unit = DT (mm, UCUM, "mm") |
|||
|
2 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm, UCUM, "mm") |
|||
|
3 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (G-A437, SRT, "Maximum") $Unit = DT (mm, UCUM, "mm") |
|||
|
4 |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-00317, SRT, "Mean") $Unit = DT (mm, UCUM, "mm") |
|||
|
5 |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-10047, SRT, "Standard Deviation") $Unit = DT (mm, UCUM, "mm") |
Content Item Descriptions
|
Row 1 |
Measured along the midline of the left and right contour. |