This macro records the result of a Luminance Uniformity QA task for a Display Subsystem.
Table C.32.3.3-1. Luminance Uniformity Result Macro Attributes
|
The number of luminance measurements recorded. The number depends on the test pattern specified in the Measurement Pattern Code Sequence (0028,702D). |
||
|
Definition of the locations on the display device at which luminance measurements were taken. One measurement is recorded for each location. |
||
|
Baseline CID 8302 “Measurement Pattern Codes”. |
||
|
The integer Digital Driving Level (DDL) at which the luminance values were measured. |
||
|
Whether white point values were also recorded at each location. |
||
|
Recorded Luminance Response Values The number of Items shall match the value of Number of Luminance Points (0028, 701B). The Items shall be stored in major row order (from left to right then from top to bottom). |
||
|
Luminance in candelas per square meter (cd/m2) at the measurement location. |
||
|
White Point at the measurement location, The white point is recorded as a CIE xy chromaticity with two dimensionless floating point values where the first is the x value and the second is the y value. |
||
|
Reflected ambient light in candelas per square meter (cd/m2) when the measurements were taken. |
||