This general Template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Table TID 10002. Accumulated X-Ray Dose
|
DTID 10004 “Accumulated Fluoroscopy and Acquisition Projection X-Ray Dose” |
IFF TID (10001) Row 4 = (113957, DCM, "Fluoroscopy-Guided Projection Radiography System") or TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") and TID (10001) Row 4 is absent) |
|||||||
|
IFF TID (10001) Row 4 = (113958, DCM, "Integrated Projection Radiography System") or TID (10001) Row 4 = (113957, DCM, "Fluoroscopy-Guided Projection Radiography System") or TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") and TID (10001) Row 4 is absent) |
||||||||
|
DTID 10006 “Accumulated Cassette-based Projection Radiography Dose” |
IFF TID (10001) Row 4 = (113959, DCM, "Cassette-based Projection Radiography System") |
|||||||
|
Required if the irradiating device is not the recording device and the dose was accumulated on a single device. |
$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device") |
|
Date that the calibration of the equipment's dose indicators was performed |
|
|
Typically a value provided by the medical physicist. The recorded dose or dose area product values in this report can be multiplied by this factor to obtain estimated real-world values. |
|
|
Value range from 0 to 100 percent. Uncertainty of the 'actual' value expressed as+/- of the mean. |
|
|
Identifies Individual or organization responsible for calibration |
|
|
Describes calibration protocol according to equipment standards or local guidelines. |
|
|
The device that produced the irradiation accumulated in this Template. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more than one device produced the accumulated irradiation. |