ECG Report Templates 

TID 3700 ECG Report

The ECG Report Template is the root structure for the representation of analysis of an ECG waveform, potentially in comparison to a prior ECG waveform analysis. The analyzed waveform may or may not be stored as a DICOM SOP Instance.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 3700. ECG Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (28010-7, LN, "ECG Report")

1

M

Root node

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

U

DCID 3670 “ECG Procedure Types”

3

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

4

>

CONTAINS

INCLUDE

DTID 1002 “Observer Context”

1-n

M

5

>

CONTAINS

CONTAINER

EV (18785-6, LN, "Indications for Procedure")

1

U

6

>>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

U

DCID 3671 “Reason for ECG Exam”

7

>>

CONTAINS

TEXT

EV (121071, DCM, "Finding")

1

U

8

>

CONTAINS

INCLUDE

DTID 3802 “Cardiovascular Patient History”

1

U

9

>

CONTAINS

INCLUDE

DTID 3704 “Patient Characteristics for ECG”

1

U

10

>

CONTAINS

INCLUDE

DTID 3702 “Prior ECG Exam”

1

U

11

>

CONTAINS

INCLUDE

DTID 3708 “ECG Waveform Information”

1

M

12

>

CONTAINS

CONTAINER

EV (122144, DCM, "Quantitative Analysis")

1

M

13

>>

CONTAINS

INCLUDE

DTID 3713 “ECG Global Measurements”

1

U

14

>>

CONTAINS

INCLUDE

DTID 3714 “ECG Lead Measurements”

1-n

U

One instantiation per reported lead

15

>

CONTAINS

INCLUDE

DTID 3717 “ECG Qualitative Analysis”

1

U

16

>

CONTAINS

INCLUDE

DTID 3719 “Summary, ECG”

1

U