Table A.75-1. Parametric Map IOD Modules
|
IE |
Module |
Reference |
Usage |
|---|---|---|---|
|
Patient |
Patient |
M |
|
|
Clinical Trial Subject |
U |
||
|
Study |
General Study |
M |
|
|
Patient Study |
U |
||
|
Clinical Trial Study |
U |
||
|
Series |
General Series |
M |
|
|
Parametric Map Series |
M |
||
|
Clinical Trial Series |
U |
||
|
Frame of Reference |
Frame of Reference |
M |
|
|
Equipment |
General Equipment |
M |
|
|
Enhanced General Equipment |
M |
||
|
Image |
General Image |
M |
|
|
Image Pixel |
C - Required if integer pixels |
||
|
Floating Point Image Pixel |
C - Required if 32 bit floating point pixels |
||
|
Double Floating Point Image Pixel |
C - Required if 64 bit floating point pixels |
||
|
Parametric Map Image |
M |
||
|
Multi-frame Functional Groups |
M |
||
|
Multi-frame Dimension |
M |
||
|
Cardiac Synchronization |
U |
||
|
Respiratory Synchronization |
U |
||
|
Bulk Motion Synchronization |
U |
||
|
Acquisition Context |
M |
||
|
Device |
U |
||
|
Specimen |
U |
||
|
Common Instance Reference |
C - Required if Referenced Image Functional Group (Section C.7.6.16.2.5) or Derivation Image Functional Group (Section C.7.6.16.2.6) is present. |
||
|
SOP Common |
M |
||
|
Frame Extraction |
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request |