Goal: A Hanging Protocol for MR & CT of Head, for a neurosurgery plan. 1Kx1K screen on left shows orthogonal MPR slices through the acquisition volume, and in one presentation group has a 3D interactive volume rendering in the lower right quadrant. In all display sets the 1Kx1K screen is split into 4 512x512 quadrants. The 2560x2048 screen has a 4 row by 3 column tiled display area. There are 4 temporal presentation groups: CTnew, MR, combined CTnew and MR, combined CTnew and CTold.
Display Environment Spatial Position attribute values for image boxes are represented in terms of ratios in pixel space [(0/3072, 512/2560), (512/3072,0/2560)] rather than (0.0,0.0), (1.0,1.0) space, for ease of understanding the example.
Hanging Protocol Name: "NeurosurgeryPlan"
Hanging Protocol Description: "Neurosurgery planning, requiring MR and CT of head"
Hanging Protocol Level: "SITE"
Hanging Protocol Creator: "Smith^Joseph"
Hanging Protocol Creation DateTime: "20020101104200"
Hanging Protocol Definition Sequence:
Item 1:
Modality: "MR"
Anatomic Region Sequence:
Item 1: (T-D1100, SRT, "Head")
Laterality: zero length
Procedure Code Sequence:
Item 1: (98765, 99Local, 1.5, "NeuroSurgery Plan Local5")
Reason for Requested Procedure Code Sequence:
Item 1: (I67.1, I10, "Cerebral aneurysm")
Item 2:
Modality: "CT"
Anatomic Region Sequence:
Item 1: (T-D1100, SRT, "Head")
Laterality: zero length
Procedure Code Sequence:
Item 1: (98765, 99Local, 1.5, "NeuroSurgery Plan Local5")
Reason for Requested Procedure Code Sequence:
Item 1: (I67.1, I10, "Cerebral aneurysm")
Number of Priors Referenced: 1
Image Sets Sequence:
Item 1:
Image Set Selector Sequence:
o Item 1:
Image Set Selector Usage Flag: "NO_MATCH"
Selector Attribute: (0018,0015) [Body Part Examined]
Selector Attribute VR: "CS"
Selector CS Value: "HEAD"
Selector Value Number: 1
Item 2:
Image Set Selector Usage Flag: "NO_MATCH"
Selector Attribute: (0008,0060) [Modality]
Selector Attribute VR: "CS"
Selector CS Value: "MR"
Selector Value Number: 1
Time Based Image Sets Sequence:
o Item 1:
Image Set Number: 1
Image Set Selector Category: "RELATIVE_TIME"
Relative Time: 0\0
Relative Time Units: "MINUTES"
Image Set Label: "Current MR Head"
Item 2:
Image Set Selector Sequence:
Item 1:
Image Set Selector Usage Flag: "NO_MATCH"
Selector Attribute: (0018,0015) [Body Part Examined]
Selector Attribute VR: "CS"
Selector CS Value: "HEAD"
Selector Value Number: 1
o Item 2:
Image Set Selector Usage Flag: "NO_MATCH"
Selector Attribute: (0008,0060) [Modality]
Selector Attribute VR: "CS"
Selector CS Value: "CT"
Selector Value Number: 1
Time Based Image Sets Sequence:
Item 1:
Image Set Number: 2
Image Set Selector Category: "RELATIVE_TIME"
Relative Time: 0\0
Relative Time Units: "MINUTES"
Image Set Label: "Current CT Head"
Item 2:
Image Set Number: 3
Image Set Selector Category: "ABSTRACT_PRIOR"
Abstract Prior Value: 1\1
Image Set Label: "Prior CT Head"
Hanging Protocol User Identification Code Sequence: zero length
Hanging Protocol User Group Name: "ABC Hospital"