Each instance of the CAD Common Image Properties Entry template contains selected attributes for a set of parallel contiguous equally spaced slices (with identical properties) from which CAD findings are derived.
Extensible
Significant
Table TID 4122. CAD Common Image Properties Entry
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
CONTAINS |
UIDREF |
1 |
M |
|||
|
3 |
> |
CONTAINS |
UIDREF |
1 |
M |
|||
|
4 |
> |
CONTAINS |
DATE |
1 |
M |
Shall be taken from Study Date (0008,0020) in the Image Instances. |
||
|
5 |
> |
CONTAINS |
TIME |
1 |
M |
Shall be taken from Study Time (0008,0030) in the Image Instances. |
||
|
6 |
> |
CONTAINS |
CODE |
1 |
M |
Shall be taken from Modality (0008,0060) in the Image Instances. |
||
|
7 |
> |
CONTAINS |
NUM |
1 |
M |
Shall be taken from value 1 of Pixel Spacing (0028,0030) in the Image Instances. UNITS = EV (mm/{pixel}, UCUM, "millimeters per pixel") |
||
|
8 |
> |
CONTAINS |
NUM |
1 |
M |
Shall be taken from value 2 of Pixel Spacing (0028,0030) in the Image Instances. UNITS = EV (mm/{pixel}, UCUM, "millimeters per pixel") |
||
|
9 |
> |
CONTAINS |
NUM |
1 |
M |
Shall be taken from Slice Thickness (0018,0050) in the Image Instances. UNITS = EV (mm, UCUM, "millimeter") |
||
|
10 |
> |
CONTAINS |
NUM |
1 |
M |
Shall be computed from the Image Position (Patient) (0020,0032) projected onto the normal to the Image Orientation (Patient) (0020,0037); may or may not be the same as the Spacing Between Slices (0018,0088) if present. UNITS = EV (mm, UCUM, "millimeter") |
||
|
11 |
> |
CONTAINS |
CODE |
EV (112228, DCM, "Recumbent Patient Position with respect to gravity") |
1 |
MC |
Required if Patient Position (0018,5100) is present in the image instances and has a value. |
Shall be derived from Patient Position (0018,5100) in the Image Instances. |