The Blood Velocity Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and containing primary (not derived) velocity measurements, e.g., from a Doppler catheter. Derived velocity measurements may be recorded using TID 3560 “Derived Hemodynamic Measurements”.
Extensible
Significant
Table TID 3509. Blood Velocity Measurement
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
INCLUDE |
1 |
M |
$LocationName = EV (G-C0E9, SRT, "Procedure site") $LocationValue = BCID 3606 “Arterial Source Locations” |
|||
|
3 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = DCID 3612 “Blood Velocity Measurements” $Units = EV (mm/s, UCUM, "mm/s") |
||
|
4 |
> |
CONTAINS |
INCLUDE |
1-n |
U |