The Angiographic Lesion Analysis Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.
Extensible
Significant
Table TID 3215. Angiographic Lesion Analysis
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
EV (F-00585, SRT, "Lesion Finding") |
1 |
M |
||||
|
2 |
> |
CONTAINS |
TEXT |
1 |
M |
|||
|
3 |
>> |
HAS PROPERTIES |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
|
4 |
>>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
|
5 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm, UCUM, "mm") |
||
|
6 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Derivation = EV (R-404FB, SRT, "Minimum") $Unit = DT (mm2, UCUM, "mm2") |
||
|
7 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
8 |
> |
CONTAINS |
CONTAINER |
1 |
U |
|||
|
9 |
>> |
CONTAINS |
NUM |
1-n |
M |
UNITS = DT (mm, UCUM, "mm") |
||
|
10 |
>>> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Unit = DT (mm, UCUM, "mm") |
||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $TargetSite = (122382, DCM, "Site of Luminal Minimum") $Unit = DT (mm, UCUM, "mm") |
||
|
12 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (122404, DCM, "Reconstructed") $TargetSite = (122382, DCM, "Site of Luminal Minimum") $Unit = DT (mm2, UCUM, "mm2") |
||
|
13 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-41D2D, SRT, "Calculated") $TargetSite = EV (122481, DCM, "Contour Start") $Unit = DT (mm, UCUM, "mm") |
||
|
14 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (G-0364, SRT, "Vessel Luminal Diameter") $Derivation = EV (R-41D2D, SRT, "Calculated") $TargetSite = EV (122482, DCM, "Contour End") $Unit = DT (mm, UCUM, "mm") |
||
|
15 |
> |
CONTAINS |
INCLUDE |
1 |
M |
|||
|
16 |
> |
CONTAINS |
CONTAINER |
EV (122517, DCM, "Densitometric Luminal Cross-sectional Area Graph") |
1 |
U |
||
|
17 |
>> |
CONTAINS |
NUM |
1 |
M |
Value = 1 UNITS = DT ({pixels}, UCUM, "pixels") |
||
|
18 |
>> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Unit = (mm2, UCUM, "mm2") |
||
|
19 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (R-41D2D, SRT, "Calculated") $Method = EV (122474, DCM, "Densitometric method") $TargetSite = EV (122481, DCM, "Contour Start") $Unit = (mm2, UCUM, "mm2") |
||
|
20 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (G-0366, SRT, "Vessel Lumen Cross-Sectional Area") $Derivation = EV (R-41D2D, SRT, "Calculated") $Method = EV (122474, DCM, "Densitometric method") $TargetSite = EV (122482, DCM, "Contour End") $Unit = (mm2, UCUM, "mm2") |
||
|
21 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (R-101BC, SRT, "Stenotic Lesion Length") $Unit = DT (mm, UCUM, "mm") |
||
|
22 |
> |
CONTAINS |
INCLUDE |
1 |
M |
$Measurement = EV (R-101BB, SRT, "Lumen Diameter Stenosis") $Unit = DT (%, UCUM, "%") |
||
|
23 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (R-101BA, SRT, "Lumen Area Stenosis") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Unit = DT (%, UCUM, "%") |
||
|
24 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$Measurement = EV (122372, DCM, "Lumen Volume") $Method = DCID 3470 “Vessel Lumen Cross-sectional Area Calculation Methods” $Unit = DT (mm3, UCUM, "mm3") |
||
|
25 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122542, DCM, "Plaque Area") $Unit = DT (mm2, UCUM, "mm2") |
||
|
26 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122376, DCM, "Total Plaque Volume") $Unit = DT (mm3, UCUM, "mm3") |
||
|
27 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122544, DCM, "Diameter Symmetry") $Unit = DT ({ratio}, UCUM, "ratio") |
||
|
28 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122545, DCM, "Area Symmetry") $Unit = DT ({ratio}, UCUM, "ratio") |
||
|
29 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122546, DCM, "Inflow Angle") $Unit = DT (deg, UCUM, "deg") |
||
|
30 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$Measurement = EV (122547, DCM, "Outflow Angle") $Unit = DT (deg, UCUM, "deg") |
||
|
31 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
32 |
> |
CONTAINS |
IMAGE |
No purpose of reference |
1 |
U |
Content Item Descriptions
|
Row 8 |
Set of user defined reference position for method that requires local reference position. |
|
Row 9 |
Distance of local reference position from an arbitrary landmark. |
|
Row 10 |
Diameter at a local reference position. |
|
Row 11 |
The reference diameter for the arterial lesion calculated with the applicable reference method |
|
Row 12 |
The reference area for the arterial lesion calculated with the applicable reference method |
|
Row 13 |
The diameter measurement at the start of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14) |
|
Row 14 |
The diameter measurement at the end of the reconstruction line in the diameter graph (TID 3214 “Analyzed Segment” Row 14) |
|
Row 15 |
The positions of the lesion, borders of the lesion, etc. |
|
Row 16 |
The graph with the calculated cross sectional area results based on the densitometric method |
|
Row 18 |
The cross sectional area measurements calculated based on the densitometric method |
|
Row 19 |
The cross sectional area measurement at the start of the reconstruction line in the area graph |
|
Row 20 |
The cross sectional area measurement at the end of the reconstruction line in the area graph |
|
Row 21 |
Measured along the midline of the left and right contour |
|
Row 22 |
The diameter stenosis is calculated as follows: (Reference Luminal Diameter - Minimum Luminal Diameter / Reference Luminal Diameter) * 100% |
|
Row 23 |
The circular and the densitometric area stenosis are calculated respectively as: (Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Circular Area / Reference Vessel Lumen Cross-Sectional Area) * 100%(Reference Vessel Lumen Cross-Sectional Area - Minimum Luminal Densitometric Area / Reference Vessel Lumen Cross-Sectional Area) * 100% |
|
Row 24 |
Estimated lumen volume between proximal border and distal border of lesion (TID 3218 “Position in Arterial Segment”, row 1 and 2) |
|
Row 32 |
Secondary Capture image with obstruction analysis contour |