This template provides information on total Projection Radiography dose values accumulated on Integrated or combined fluoroscopy/acquisition systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.
Extensible
Non-Significant
Table TID 10007. Accumulated Total Projection Radiography Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1 |
M |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
||||
|
2 |
NUM |
1 |
MC |
IF TID (10001) Row 4 = (113958, DCM, "Integrated Projection Radiography System") or any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). |
UNITS = EV (Gy, UCUM, "Gy") |
|||
|
3 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
4 |
NUM |
1 |
U |
UNITS = EV (1, UCUM, "no units") |
||||
|
5 |
CODE |
1 |
MC |
IF any of (113725, DCM, "Dose (RP) Total"), (113728, DCM, "Fluoro Dose (RP) Total") or (113729, DCM, "Acquisition Dose (RP) Total") are present, and Row 6 is not present. |
||||
|
6 |
TEXT |
1 |
MC |
IF any of (113725, DCM, "Dose (RP) Total"), (113728, DCM, "Fluoro Dose (RP) Total") or (113729, DCM, "Acquisition Dose (RP) Total") are present, and Row 5 is not present. |
Content Item Descriptions
|
Row 1 |
Accumulated Dose Area Product |
|
Row 2 |
Accumulated dose relative to reference point. |
|
Row 3 |
A single value for Radiography systems calculating reference point dose based on fixed distance. |
|
Row 5 |
A coded definition of the Reference Point (RP) used for RP-related dose values. |
|
Row 6 |
A text definition of the Reference Point (RP) used for RP-related dose values. |