This Module describes the attributes related to dose delivery from an X-Ray source during the acquisition of an X-Ray image.
Table C.8-33. X-Ray Acquisition Dose Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
KVP |
(0018,0060) |
3 |
Peak kilo voltage output of the X-Ray generator used. |
|
X-Ray Tube Current |
(0018,1151) |
3 |
X-Ray Tube Current in mA. |
|
X-Ray Tube Current in µA |
(0018,8151) |
3 |
X-Ray Tube Current in µA. |
|
Exposure Time |
(0018,1150) |
3 |
Duration of X-Ray exposure in msec. |
|
Exposure Time in µS |
(0018,8150) |
3 |
Duration of X-Ray exposure in µsec. |
|
Exposure |
(0018,1152) |
3 |
The exposure expressed in mAs, for example calculated from Exposure Time and X-Ray Tube Current. |
|
Exposure in µAs |
(0018,1153) |
3 |
The exposure expressed in µAs, for example calculated from Exposure Time and X-Ray Tube Current. |
|
Distance Source to Detector |
(0018,1110) |
3 |
Distance in mm from source to detector center. NoteThis value is traditionally referred to as Source Image Receptor Distance (SID). |
|
Distance Source to Patient |
(0018,1111) |
3 |
Distance in mm from source to the table, support or bucky side that is closest to the Imaging Subject, as measured along the central ray of the X-Ray beam. Note
|
|
Image and Fluoroscopy Area Dose Product |
(0018,115E) |
3 |
X-Ray dose, measured in dGy*cm*cm, to which the patient was exposed for the acquisition of this image plus any non-digitally recorded fluoroscopy that may have been performed to prepare for the acquisition of this image. Note
|
|
Body Part Thickness |
(0018,11A0) |
3 |
The average thickness in mm of the body part examined when compressed, if compression has been applied during exposure. |
|
Relative X-Ray Exposure |
(0018,1405) |
3 |
Indication of the applied dose, in manufacturer specific units. Note
|
|
Entrance Dose |
(0040,0302) |
3 |
Average entrance dose value measured in dGy at the surface of the patient during the acquisition of this image. NoteThis may be an estimated value based on assumptions about the patient's body size and habitus. |
|
Entrance Dose in mGy |
(0040,8302) |
3 |
Average entrance dose value measured in mGy at the surface of the patient during the acquisition of this image. NoteThis may be an estimated value based on assumptions about the patient's body size and habitus. |
|
Exposed Area |
(0040,0303) |
3 |
Typical dimension of the exposed area at the detector plane. If Rectangular: row dimension followed by column; if Round: diameter. Measured in cm. Note
|
|
Distance Source to Entrance |
(0040,0306) |
3 |
Distance in mm from the source to the surface of the patient closest to the source during the acquisition of this image. NoteThis may be an estimated value based on assumptions about the patient's body size and habitus. |
|
Comments on Radiation Dose |
(0040,0310) |
3 |
User-defined comments on any special conditions related to radiation dose encountered during the acquisition of this image. |
|
X-Ray Output |
(0040,0312) |
3 |
The X-Ray output at the patient entrance surface and kVp used to acquire the image, measured in mGy/mAs. NoteThis value may be a calibrated value rather than measured during the exposure. |
|
Half Value Layer |
(0040,0314) |
3 |
The thickness of Aluminum in mm required to reduce the X-Ray Output (0040,0312) by a factor of two. NoteThis value may be a calibrated value rather than measured during the exposure. |
|
Organ Dose |
(0040,0316) |
3 |
Average organ dose value measured in dGy during the acquisition of this image. NoteThis may be an estimated value. |
|
Organ Exposed |
(0040,0318) |
3 |
Organ to which Organ Dose (0040,0316) applies. Defined Terms:
NoteThe anatomic regions described by these terms are those that are particularly radiosensitive and for which it is conventional to obtain organ specific dose parameters. |
|
Anode Target Material |
(0018,1191) |
3 |
The primary material in the anode of the X-Ray source. Defined Terms:
|
|
Rectification Type |
(0018,1156) |
3 |
Type of rectification used in the X-Ray generator. Defined Terms:
|