Table A.54-1. X-Ray 3D Craniofacial Image IOD Modules
|
IE |
Module |
Reference |
Usage |
|---|---|---|---|
|
Patient |
Patient |
M |
|
|
Clinical Trial Subject |
U |
||
|
Study |
General Study |
M |
|
|
Patient Study |
U |
||
|
Clinical Trial Study |
U |
||
|
Series |
General Series |
M |
|
|
Clinical Trial Series |
U |
||
|
Enhanced Series |
M |
||
|
Frame of Reference |
Frame of Reference |
M |
|
|
Equipment |
General Equipment |
M |
|
|
Enhanced General Equipment |
M |
||
|
Image |
Image Pixel |
M |
|
|
Enhanced Contrast/Bolus |
C - Required if contrast media was applied. |
||
|
Device |
U |
||
|
Intervention |
U |
||
|
Acquisition Context |
M |
||
|
Multi-frame Functional Groups |
M |
||
|
Multi-frame Dimension |
U |
||
|
Patient Orientation |
U |
||
|
Image - Equipment Coordinate Relationship |
U |
||
|
Specimen |
U |
||
|
X-Ray 3D Image |
M |
||
|
X-Ray 3D Craniofacial Image Contributing Sources |
U |
||
|
X-Ray 3D Craniofacial Acquisition |
U |
||
|
X-Ray 3D Reconstruction |
U |
||
|
SOP Common |
M |
||
|
Common Instance Reference |
U |
||
|
Frame Extraction |
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request |
The Overlay Plane Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the X-Ray 3D Craniofacial Image.
The VOI LUT function is provided by a Frame VOI LUT Functional Group.
The Curve Module was previously included in the Image IE for this IOD but has been retired. See PS3.3-2004.