This Module describes the attributes related to generation of X-Rays during the acquisition of an X-Ray image.
Table C.8-34. X-Ray Generation Module Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
KVP |
(0018,0060) |
3 |
Peak kilo voltage output of the X-Ray generator used. |
|
X-Ray Tube Current |
(0018,1151) |
3 |
X-Ray Tube Current in mA. |
|
X-Ray Tube Current in µA |
(0018,8151) |
3 |
X-Ray Tube Current in µA. |
|
Exposure Time |
(0018,1150) |
3 |
Duration of X-Ray exposure in msec. |
|
Exposure Time in µS |
(0018,8150) |
3 |
Duration of X-Ray exposure in µsec. |
|
Exposure |
(0018,1152) |
3 |
The exposure expressed in mAs, for example calculated from Exposure Time and X-Ray Tube Current. |
|
Exposure in µAs |
(0018,1153) |
3 |
The exposure expressed in µAs, for example calculated from Exposure Time and X-Ray Tube Current. |
|
Exposure Control Mode |
(0018,7060) |
3 |
Type of exposure control. Defined Terms:
|
|
Exposure Control Mode Description |
(0018,7062) |
3 |
Text description of the mechanism of exposure control. May describe the number and type of exposure sensors or position of the sensitive area of the imaging detector. |
|
Exposure Status |
(0018,7064) |
3 |
Whether the exposure was normally completed or not. Defined Terms:
|
|
Phototimer Setting |
(0018,7065) |
3 |
Nominal percentage phototimer setting, where a more positive value indicates greater exposure and a more negative value indicates less exposure. |
|
Focal Spot(s) |
(0018,1190) |
3 |
Nominal focal spot size in mm used to acquire this image. |
|
Anode Target Material |
(0018,1191) |
3 |
The primary material in the anode of the X-Ray source. Defined Terms:
|
|
Rectification Type |
(0018,1156) |
3 |
Type of rectification used in the X-Ray generator. Defined Terms:
|
|
Generator ID |
(0018,1005) |
3 |
Identifier of the generator |