Table A.32.8-1. VL Whole Slide Microscopy Image IOD Modules
|
IE |
Module |
Reference |
Usage |
|---|---|---|---|
|
Patient |
Patient |
M |
|
|
Clinical Trial Subject |
U |
||
|
Study |
General Study |
M |
|
|
Patient Study |
U |
||
|
Clinical Trial Study |
U |
||
|
Series |
General Series |
M |
|
|
Whole Slide Microscopy Series |
M |
||
|
Clinical Trial Series |
U |
||
|
Frame of Reference |
Frame of Reference |
M |
|
|
Equipment |
General Equipment |
M |
|
|
Enhanced General Equipment |
M |
||
|
Image |
General Image |
M |
|
|
Image Pixel |
M |
||
|
Acquisition Context |
M |
||
|
Multi-frame Functional Groups |
M |
||
|
Multi-frame Dimension |
M |
||
|
Specimen |
M |
||
|
Whole Slide Microscopy Image |
M |
||
|
Optical Path |
M |
||
|
Multi-Resolution Navigation |
C - Required if Image Type (0008,0008) Value 3 is LOCALIZER |
||
|
Slide Label |
C - Required if Image Type (0008,0008) Value 3 is LABEL; may be present otherwise |
||
|
SOP Common |
M |
||
|
Common Instance Reference |
M |
||
|
Frame Extraction |
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request |
The Code Sequences within the Optical Path Sequence (0048,0105) of the Optical Path Module (see Section C.8.12.5) are constrained as follows:
Baseline CID for Illuminator Type Code Sequence (0048,0100) is CID 8125 “Microscopy Illuminator Type”.
Baseline CID for Illumination Color Code Sequence (0048,0108) is CID 8122 “Microscopy Illuminator and Sensor Color”.
Baseline CID for Illumination Type Code Sequence (0022,0016) is CID 8123 “Microscopy Illumination Method”.
Baseline CID for Lenses Code Sequence (0022,0019) is CID 8121 “Microscopy Lens Type”.
Baseline CID for Light Path Filter Type Stack Code Sequence (0022,0017) and for Image Path Filter Type Stack Code Sequence (0022,0018) is CID 8124 “Microscopy Filter”.
Baseline CID for Channel Description Code Sequence (0022,001A) is CID 8122 “Microscopy Illuminator and Sensor Color”.